6-4
10083002
Operator Manual
Cleaning the Equipment
6.3 Areas to Be
Cleaned Before Each
Use
NOTE: Refer to the user manual supplied with the monitor for the
manufacturer's recommended procedures for cleaning the monitor
and the monitor control surface.
The following areas of the lighting system must be cleaned and
disinfected before each use of the system (see
for instructions):
•
Lighthead and Monitor Suspension Arms –
Wipe the entire
suspension arm, including the suspension fork and yoke.
•
Lighthead
– Wipe both top and side surfaces.
•
Sterile Handle Support
– Wipe all areas of the support,
including those covered when the handle is installed.
•
Lens:
See isntructions below.
Important:
Clean only the exterior surface of the lens.
1.
Remove the lighthandle.
2.
Clean/disinfect the outer surface of the lens as outlined in
3.
Wipe the outer surface of the lens with an anti-static cleaner and
soft cloth.
4.
Do not reinstall a lighthandle until immediately before the lighting
system is to be used in a surgical procedure.
•
When using a sterilizable lighthandle cover, verify that it is
sterilized according to facility sterile protocols between each
surgical procedure.
•
When using disposable sterile lighthandle covers, verify a
new one is used for each procedure.
NOTE: Always sterilize metal sterilizable handle between
procedures. Sterilize lighthandle using conventional hospital
sterilization procedures and a standard sterilization
prevacuum or gravity cycle.
•
Control Center –
Wipe housing and control center with an anti-
static cleaner and soft cloth.
WARNING – PERSONAL
INJURY HAZARD: Do not
attempt to clean lighthead
unless power is turned off and
the lighthead has cooled
sufficiently.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE
HAZARD:
•
Cleaning and disinfecting
agents used on this lighting
system must be certified by
their manufacturer to be
compatible with the following
materials: polycarbonate,
polyetherimide, santoprene.
•
DO NOT SPRAY any cleaning
product directly onto the
lighthead, modem chassis, or
any system components.
•
Use only recommended
cleaning/disinfecting and/or
anti-static agents on this light.
Some degree of staining,
pitting, and/or discoloration
could occur if a phenolic-,
iodophor-, or glutaraldehyde-
based disinfectant is used on
the surfaces of this light. Also,
use of alcohol or aerosol spray
cleaner/disinfectants
containing a substantial
amount of alcohol in the
formula can damage the
polycarbonate lens.
•
Do not scratch optical coating
on accessible portions of optic
assembly when cleaning;
always wear rubber gloves and
use only a clean, white, lint-
free cloth when wiping
external surfaces.
•
Prevent leakage of fluids into
interior of lighthead or control
center.
!
!
Содержание Harmony LED Wall Control Unit
Страница 2: ......
Страница 8: ...vi 10083002 Operator Manual Table of Contents Healthcare Capital Equipment ...
Страница 10: ...viii 10083002 Operator Manual Table of Contents Healthcare Capital Equipment ...
Страница 18: ...2 4 10083002 Operator Manual Symbol Definitions Healthcare Capital Equipment ...
Страница 66: ...8 2 10083002 Operator Manual Maintenance Healthcare Capital Equipment ...
Страница 68: ...9 2 10083002 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Страница 70: ...10 2 10083002 Operator Manual Waste Management Guidelines Healthcare Capital Equipment ...