5-5
Operating Instructions
Operator Manual
10083002
If an indication of LED module or lamp failure appears on the display
(or at the lighthead, see
Section 5.3, Lighthead Fault Indications,
), refer module replacement to a STERIS-trained service
technician.
5.4.2 Control Center Fault
Log Indications
The Harmony LED Wall Control Unit is capable of displaying fault
messages. Fault messages, for the most part, are intended for use
by trained STERIS service technicians. An operator not trained in
servicing this equipment can provide fault message information to
STERIS when communicating a service issue. A summary of faults
that can appear in the fault log is given in
Important:
Refer any problems with “yes” listed under the
Service
Required
column to STERIS. Also, “#” indicates a numeric value on
the display.
Table 5-2. Control Center Fault Indications
Fault Message
Description
Service
Required
POWER SUPPLY OR MAINS
FAILURE LIGHTHEAD: #
System alternating current (ac) power supply failure on a
lighthead. LED module failure indicator flashes on the Light
Control Screen.
yes
COMMUNICATION FAILURE AT
LIGHTHEAD: #
The control center can no longer communicate with a
lighthead.
yes
BATTERY POWER SUPPLIED TO
LIGHTHEAD: #
Displayed when one or more lightheads are operating on
battery backup power.
no
CAMERA COMMUNICATION
FAILURE ON LIGHTHEAD: #
The lighthead cannot communicate with camera.
yes
CAMERA MOTOR ROTATION -
OVER CURRENT ON
LIGHTHEAD: #
The motor that rotates the camera is drawing excessive
current (camera is not rotating smoothly or is binding inside
the camera module assembly).
yes
SWITCH FAILURE ON
LIGHTHEAD: #
Pressure on a lighthead button has been detected for
40 seconds. Switch is presumed to be stuck.
yes
SWITCH FAILURE ON FRONT
PANEL - SW NO: #
Pressure on a control center front panel key has been
detected for 40 seconds. Switch is presumed to be stuck.
yes
MULTIPLE CAMERAS DETECTED
ON LIGHTHEADS: #
More than one lighthead is reporting that it has a camera
module installed.
yes
MAIN CONTROLLER
COMMUNICATION FAILURE
Control center main controller and display controllers cannot
communicate with each other.
yes
INVALID ASSEMBLY ID ON
LIGHTHEAD:
The lighthead firmware does not recognize the configuration
of its lighthead board.
yes
LED MODULES N1, N2 FAILURE
ON LIGHTHEAD: #
Low current through one or more LED Module drivers.
yes
LED MODULES N1, N2 OVER-
CURRENT FAILURE ON
LIGHTHEAD: #
High current through one or more LED Module drivers.
yes
Содержание Harmony LED Wall Control Unit
Страница 2: ......
Страница 8: ...vi 10083002 Operator Manual Table of Contents Healthcare Capital Equipment ...
Страница 10: ...viii 10083002 Operator Manual Table of Contents Healthcare Capital Equipment ...
Страница 18: ...2 4 10083002 Operator Manual Symbol Definitions Healthcare Capital Equipment ...
Страница 66: ...8 2 10083002 Operator Manual Maintenance Healthcare Capital Equipment ...
Страница 68: ...9 2 10083002 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Страница 70: ...10 2 10083002 Operator Manual Waste Management Guidelines Healthcare Capital Equipment ...