7-13
Troubleshooting
Operator Manual
920017-148
Open or short circuit on Pressure Transmitter.
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Call STERIS.*
29.
If Drying System option is present,
alarm
sounds and screen displays:
ALARM: DRYING
TOO LONG TO HEAT
Printer records:
* ALARM
DRYING
TOO LONG TO HEAT
HH:MM:SSX
YY/MM/DD
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Drying temperature time-out alarm has elapsed before air
temperature reaches setpoint – Adjust time-out alarm delay in
Service Mode.
3. Steam supply valves closed – Open steam supply valves.
4. Steam pressure below specifications — Call STERIS.*
28. Alarm sounds and screen displays:
ALARM: PUMP PRESSURE
IS DEFECTIVE
Printer records:
* ALARM
PUMP PRESSURE
IS DEFECTIVE
HH:MM:SSX
YY/MM/DD
Table 7-2. Troubleshooting Chart – Alarm Screen and/or Printout (Cont'd)
PROBLEM
POSSIBLE CAUSE AND/OR CORRECTION
*Service charges may be incurred. Consult your warranty for details.
30. Water flowmeter is defective. Alarm sounds
and screen displays:
ALARM: WATER FLOW
METER IS DEFECTIVE
Printer records:
* ALARM
WATER FLOW
METER IS DEFECTIVE
HH:MM:SSX
YY/MM/DD
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Call STERIS.*
31. Alarm sounds and screen displays:
ALARM: ALKALINE
PUMP IS DEFECTIVE
Printer records:
* ALARM
ALKALINE
PUMP IS DEFECTIVE
HH:MM:SSX
YY/MM/DD
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Alkaline chemical injection pump still energized even if control
de-energized pump — Call STERIS.*