33
8.4
Mechanical Performance Issues
PROBLEM
POTENTIAL CAUSE
SOLUTION
Barcode scanning fails
Barcode scanner cannot read
barcode label on vial.
Check the barcode label is not
damaged. Twist the vial and scan
the barcode again.
Unsuccessful tip pickup
Tip rack not secure or seated
properly in the carousel.
Ensure tip rack is properly inserted
and flush against the carousel.
RoboSep™-S arm disturbed during a
protocol run.
Take note of the step that run(s) is
at. Abort current run. If appropriate,
finish separation manually,
otherwise prepare new sample and
restart the protocol from the
beginning. If issue persists, contact
Technical Support.
Tips not well aligned with tubes
Tip rack not secure or seated
properly in the carousel.
Ensure tip rack is properly inserted
and flush against the carousel.
RoboSep™-S arm disturbed during a
protocol run.
Take note of the step that run(s) is
at. Abort current run. If appropriate,
finish separation manually,
otherwise prepare new sample and
restart the protocol from the
beginning. If issue persists, contact
Technical Support.
Tip stripping failure
Plastic buildup on tip head.
Clean tip head using RoboSep™
Tip Head Polishing Compound
(section 6.1). Do not use ethanol.
Robotic arm stuck and bouncing up
and down in fully extended position
Homing flag on carousel is
misaligned.
Power off the instrument and
physically move the arm all the way
down. Start instrument again. If
issue persists, contact Technical
Support.
Tip rack separates when removing
the lid to open it
Excess force applied when removing
the tip rack lid.
Ensure only the top lid is removed.
Instrument waits for approximately
one minute with tip attached
This is a normal operation to
optimize the overall run time of the
instrument.
Contact STEMCELL Technical
Support to change the default
threshold time.