STEINDORFF

®

 

 

8.

 

Vertical Illumination Alignment 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.

 

Field Diaphragm Adjustment Knob 

2.

 

Aperture Diaphragm Adjustment Knob 

3.

 

Filter Wheel 

4.

 

Condenser Adjustment Knob 

5.

 

Lamp Vertical Adjustment Knob 

6.

 

Lamp Transverse Adjustment Knob 

7.

 

Field Diaphragm Centering Screw 

8.

 

Polarizer 

 

1.

 

The aperture diaphragm near the lamp house may be adjusted slightly to change contrast. 

2.

 

To center the incident or vertical lamp: 

a)

 

Put a piece of paper and place it on the stage. Take out one objective 

form nosepiece and rotate the nosepiece until this objective opening hole 

is in position over the paper.  

b)

 

Turn the power on and adjust the brightness control to establish sufficient 

light. 

c)

 

This paper will allow for focusing of the lamp filament on it. If the 

filament image is not centered in the overpass hole, reduce the aperture 

Filament  Image

 

Overpass Hole

 

Figure 10

 

 

Содержание NYMCS-1500

Страница 1: ...nt images please make a few trial snaps making sure you can operate the microscope properly 3 LCD screen was made by precision technology 99 99 of its pixels are up to specifications 0 01 of the pixel...

Страница 2: ...Make sure the table is smooth horizontal and firm enough 3 When moving the instrument please use two hands to grip with the two sides of the microscope body 4 If the bacterium solution or the water s...

Страница 3: ...Holder 7 Condenser with Aperture Diaphragm 8 Polarizer 9 Collector 10 Power Switch 11 Transmitted Reflection Illumination Toggle Switch 12 Fine Focus Control Knob 13 Coarse Focus Control 14 Focus Tens...

Страница 4: ...h 25mm Field 18mm Stage Circular rotating stage Diameter 150mm 360 graduated in 1 increments minimum retardation resolution 6 center adjustable and locking clamp for stage rotation Condenser Strain fr...

Страница 5: ...ected and the fuse holder located on the power Inlet 1 2 Press the power switch on the side of I which means the circuits is got through If the light does not appear to be ON check the brightness cont...

Страница 6: ...ntact the stage or slide glass when using the microscope Its use also simplifies focusing After focusing on the slide glass with the coarse adjustment rotation of the lever toward the rear of the micr...

Страница 7: ...denser 5 2 Condenser Alignment to center the abbe condenser in the optical light path See Figure 8 a Turn the 10X objective to optics path close iris diaphragm and lower the Abbe condenser until obser...

Страница 8: ...ensator to light path 6 3 The push rod to the upper left of the trinocular head will send 100 of the image light to the binocular eyepieces or to the top port It satisfy high grape photography NOTE If...

Страница 9: ...e adjusted slightly to change contrast 2 To center the incident or vertical lamp a Put a piece of paper and place it on the stage Take out one objective form nosepiece and rotate the nosepiece until t...

Страница 10: ...r and ground glass you may turn anyone to the light path to obtain an image of good quality 3 2 Changing the Light Bulb and the Fuse 1 Changing the lamp of the vertical illuminator See figure a Discon...

Страница 11: ...nstalling the new lamp be careful not to touch the glass with your fingers The new lamp should be supplied in a plastic protective envelope If not use a tissue or other medium to grasp the lamp This w...

Страница 12: ...ons 1 Crosshair Coordinate 2 Color White Black Red Purple Green 3 Direction key 4 Menu Setting Exit 5 Display button Only display image and remove kinds of characters symbols 6 Confirm Button 7 Snapsh...

Страница 13: ...r 3 seconds the microscope system will be in the preview in the real time automatically when you can snap C Adjust microscope s definition according to its imaging focal length make images sharp from...

Страница 14: ...tion key allows you to choose the following functions A Resolution B Camera function setting C Display setting On the snap model press menu button to enter menu interface Use direction key to select a...

Страница 15: ...you press MENU button again the model comes back to snap model 2 Video Resolution Select video resolution On video model press MENU button as shown the Resolution 1080P30 1080P30 means the pixels has...

Страница 16: ...own key to select as Scene WB DE Contrast Sharpness AE Meter AE ISO AE Bias Fig 29 30 For example Fig 29 you can press the OK key to enter the Scene setting and choose by up down keys Fig 31 Press OK...

Страница 17: ...Fig 33 You can press the OK key to enter the Date Time setting and change by direction keys Fig 34 Press OK button again to save your selection or exit by MENU button Once confirmed the model comes ba...

Страница 18: ...17 STEINDORFF B WB White Balance WB will help you to obtain superb color when observing C DE Color effects D Sharpness SET EXIT SET SET EXIT EXIT...

Страница 19: ...18 STEINDORFF E AE Meter F AE ISO G AE Bias SET EXIT SET EXIT SET EXIT...

Страница 20: ...ou can save the time B Display Date and time information can be stored together with photo taken in SD card File names also contain date and time information Before using this camera please set the da...

Страница 21: ...he SD card inserted Attention Even the protected content will be deleted when formatting cannot be resumed again Select No and press menu button the model comes back to the snap mode D Reset system se...

Страница 22: ...21 STEINDORFF F Language We developed 9 languages according to your demand G Beep H USB Selection SET EXIT SET SET EXIT EXIT...

Страница 23: ...een will show Card full B Record When you turn on the microscope you can press button to record When the card is full the LCD screen will show Card full C Playback Press button to select playback mode...

Страница 24: ...th aviation gasoline is recommended Do not use organic solvents such as alcohol ether or other thinner etc for cleaning the pointed parts or plastic components 3 Avoid disassembling the microscope Bei...

Отзывы: