25
26
Item
Operations
Display
Notes
Turn on the scale, long press
and quickly
press
two times, and enter to
calibrate
SETUP
Position of decimal
point
Press
to choose, press
to confirm
<Pot>
0: XXXXX
1: XXXX,X
2: XXX,XX
3: XX,XXX
Capacity
Press
to add unit one, press <
to move left one, press
to confirm it
<3000>
Division
press
to choose, press
to confirm.
<FD>
E.g.: 50g
,choose
“5”, 100g
choose ”1”
No need to set. Enter next step by
pressing
CAL
Auto enter to next step, no need to
set
<26000>
Input the value of standard weight
which was loaded before (e.g. 100kg)
by pressing to add 1, and pressing to
move the digit to the left. After input-
ting the value, press to confirm
<25.5>
End of calibration
<Done>
5. CALIBRATION
applicable to scale models SBS-KW-300/100-0 | SBS-KW-300/100
and SBS-KW-1TC
Symptom
Possible Causes
Solution
blank display when
On/Off
is pressed
Discharged batteries
replace batteries
press ON/OFF key for
long
Defective batteries
Defective ON/OFF key
no action taken
after
TARE or HOLD is
pressed
Defective TARE or HOLD
key
clean TARE or HOLD key
unstable readings
Scale in motion
stabilize the load and scale
Scale is damped
dry the scale
Dust on PCB board
clean PCB board
reading is not zero
without
load
Unstable system power
longer warm-up time
Load-cell stressed too
much or too long
hang the scale in storage
large error in
weight
reading
Scale is not zeroed before
loading
keep the scale unload and
reboot
Re-calibration needed
re-calibrate the scale
Improper unit
switch to proper unit
6. TROUBLE-SHOOTING
EN
EN