Steelcase 7 series Скачать руководство пользователя страница 2

READ THE ASSEMBLY DIRECTIONS AND USER GUIDE

Many times, problems can be the result of improper assembly. Reference the Assembly Directions  
document to ensure the Table is assembled correctly. If so, reference the troubleshooting steps throughout 
this document. 

TROUBLESHOOTING PARTS KIT

Having a spare part that is known to work will quickly lead to accurate identification of faulty components. 
Without this, it can be little more than guesswork.
Steelcase strongly recommends that every Dealer maintain a small kit of parts for troubleshooting purposes.

This kit of parts should include: 

Table 

1.  Lifting Column 

qty=2  Service Part no.   003072DSR 

2.  Control Box 

qty=1  Service Part no.   003071DSR 

3.  Controller (digital) 

qty=1  Service Part no.   005291DSR 

4.  Low-Voltage Cable (2m) 

qty=2  Service Part no.   003067DSR 

5.  Power Cable 

qty=1  Service Part no.   003068DSR 

Always maintain a complete kit of troubleshooting spare parts that are known to work properly, to quickly 
identify the root cause of problems and resolve customer issues on the first try.

TROUBLESHOOTING STEPS

Step 1: 

 

Check for power at the wall outlet; ensure the Table is properly assembled, and all cables are  
securely connected. 

Step 2: 

 Check all connections of Low-Voltage Cables: are any of the pins in the connectors damaged or 
not making contact? 

(see Fig. 2 on page 5)

Step 3: 

 Obtain a Control Box, Controller, Power Cable and Low-Voltage Cable that are known to work.

Step 4: 

 Using the known good components, test the Lifting Columns one by one. 

(see Common 

Procedures)

 Replace any faulty Lifting Columns.

 Between each of the following steps, be sure to initialize the Control Box. 

(see Common Procedures)

Step 5:

 

If Lifting Columns operate properly, progressively swap out parts in the test setup as follows:
A.  Swap the known-good Power Cable with the original from the non-functioning Table.  

If it stops working, the Power Cable is faulty.

B.  Using the original Power Cable, swap the known-good, Low-Voltage Cable with the original 

from the non-functioning Table. If it stops working, the Low-Voltage Cable is faulty.

C.  Using the original Power and Low-Voltage Cables, swap the known-good Controller with  

the original from the non-functioning Table. If it stops working, the Controller is faulty.

D.  Using the original Power and Low-Voltage Cables, and the Controller, swap the known-good 

Control Box with the original from the non-functioning Table. If it stops working, the Control 
Box is faulty.

Be sure to go through all of the steps above to fully identify all faulty components.  
There could be more than one!

IMPORTANT NOTE:

 This guide applies to models shipped prior to December 2015 only.

2

Содержание 7 series

Страница 1: ...Table motion and synchronizes motion of the individual Lifting Columns The Control Box will shut down the entire table if a fault is detected For example if one Lifting Column is binding or trapped it will draw significantly more current than the other Lifting Column and the Control Box will shut down the entire Table to prevent further damage The Controller is the user interface to the Table and ...

Страница 2: ...w Voltage Cables are any of the pins in the connectors damaged or not making contact see Fig 2 on page 5 Step 3 Obtain a Control Box Controller Power Cable and Low Voltage Cable that are known to work Step 4 Using the known good components test the Lifting Columns one by one see Common Procedures Replace any faulty Lifting Columns Between each of the following steps be sure to initialize the Contr...

Страница 3: ...via the software in the Control Box Operate the Table all the way up and down to confirm proper function TEST INDIVIDUAL LIFTING COLUMNS Lifting Columns can be operated individually even if they are not assembled to a table Obtain a Control Box that is known to work Be sure the Control Box is unplugged or plugged in but in Standby mode Connect the Low Voltage Cable of the targeted Lifting Column t...

Страница 4: ...inside multi pin connectors Low Voltage Cable connector Ensure the table is in Standby mode 10 seconds of inactivity Unplug each multi pin connection point and inspect the pins Are they all straight and making good contact If not try using a pushpin to straighten them see Fig 2 on page 5 Otherwise replace necessary components with undamaged versions Damaged Lifting Column wiring harness Carefully ...

Страница 5: ... Because another Lifting Column may have the same problem at a different height Lifting Columns are not parallel with one another causing them to bind First ensure that glides are properly adjusted and the Table is level Loosen but do not remove all of the screws attaching the Lifting Columns to the worksurface For Series 7 Tables ensure Stretcher is fully seated in the backs of the Lifting Column...

Страница 6: ...sed One or more buttons are pressed on passive matrix Multiple buttons pressed on 2 different handsets Check Controller E12 Cable orientation One Low Voltage Cable is not mounted correctly Low Voltage Cable plugged upside down Bad Low Voltage Cable Check Low Voltage Cable connections Replace Low Voltage Cable E13 Position error One Port Channel has a different position than others Too much back dr...

Страница 7: ...rt Channel 1 2 3 4 respectively Hit obstruction Remove obstruction Initialize the Control Box if necessary see Common Procedures E40 E41 E42 E43 Port Channel 1 2 3 4 SLS respectively Safety limit switch activated on Port Channel 1 2 3 4 respectively Hit obstruction Remove obstruction Initialize the Control Box if necessary see Common Procedures E44 E45 E46 E47 Port Channel 1 2 3 4 direction respec...

Отзывы: