STATUS SCIENTIFIC CONTROLS
Installation, Commissioning & Routine Gas Testing
FGD10B Pellistor Gas Detector
TD18/008
Issue:
10
Change Note:
1904
Page 27
4.6.9.
E : 19
– Positive Zero Suppression
This option is used to allow the user to suppress small amounts of positive
sensor zero drift. The setting can be set between 0 and 10% of the sensor
range as set by the FSD value.
•
Press MENU to open the menu system.
•
Using the
or
buttons, select menu option:
E:19
•
Press
.
•
Using the
or
buttons, set the required zero suppression
value.
•
Press ENTER to store the new value.
Note: Pressing the MENU button rather than the
R button
exits without any change.
•
Press MENU to close the menu system.
4.6.10. E : 20
– Negative Zero Suppression
This option is used to allow the user to suppress small amounts of negative
sensor zero drift. The setting can be set between 0 and -10% of the sensor
range as set by the FSD value.
•
Press MENU to open the menu system.
•
Using the
or
buttons, select menu option:
E:20
•
Press
.
•
Using the
or
buttons, set the required zero suppression
value.
•
Press
to store the new value.
Note: Pressing the MENU button rather than the
button
exits without any change.
•
Press MENU to close the menu system.