STATUS SCIENTIFIC CONTROLS
Installation, Commissioning & Routine Gas Testing
FGD10B Pellistor Gas Detector
TD18/008
Issue:
10
Change Note:
1904
Page 22
4.6.
Menu Mode Selection
The external Calibration / Configuration Keypad is used in its simplest form to calibrate
the sensor for zero or gas drift.
It may also be used to configure the FGD10b.
The following features are available via the Calibration / Configuration Keypad menu
system: -
Menu Option
E : 3
– Sensor FSD
E : 9
– Diagnostics
E : 12
– Cross reference
E : 19
– Positive zero suppression
E : 20
– Negative zero suppression
E : 25
– Zero temperature compenve
E : 34
– Span temperature compenve
E : 35
– Span temperature compensation -ve
E : 36
– Zero temperature compensation -ve
E : 77
– Firmware version
The menu options are selected as follows:
•
Press the MENU button,
E: 1
appears on the display.
•
Press
or
until the required menu option is displayed, see
following options.
•
Press
to select the menu option.
•
To exit the menu press MENU.
While the instrument is in menu mode
– any data displayed on the screen will alternate
between the menu number and the reading.