18
Status
Pro
–
µLevel User Guide
ADJUSTMENT OF THE APPARATUS
5. Keep the position of digital level meter and wait the reading to be steady, and press
and
separately, the displayer level meter show zero, then replace by a
relevant gauge listed as below under the probe of the right side indication needle,
adjust the small angle inspection instrument and make the displayed value of the indi
cation needle to zero position, at such time press the key “Save” of the control box ,
the level meter will show 500.
6. Repeat step 4
7. Set zero again to the digital level meter, then replace by a relevant gauge as above
under the probe of the left side indication needle, adjust the small angle inspection
instrument and make the displayed value of the indication needle to zero position, at
such time press the key “Delete” of the control box, the level meter will show -500.
8. Insert the instrument screwdriver into the small hole at left side of the digital level
meter, press the scale/restoration key twice, when LED show
, then it comes into the
measuring mode.
5.4 Renew the Leave Factory Settings
During the use of the product, if scale the level meter by error operate, the function of
renew the leave factory setting could help. Insert the instrument screwdriver into the
small hole at the left side of the digital level meter, press the scale/restoration key twice,
after LED show
, tough key
gently, then press again scale/restoration, after LED
show
, it comes into the measuring mode.