Function Choice
After the apparatus is switched on, touch gently Mode key
or
could choose
different functions.
Measuring Phase
This is a high precision measuring phase. The resolution of measur-
ing phase II of WL11 Digital Level Meter is 0.001mm/m, suit to
inspect the flatness, straightness of products with precision grade “00”. The resolution of
measuring phase II of WL10 Digital Level Meter is 0.005mm/m, suit to inspect the flatness,
straightness of products with precision grade “0”.
Measuring Phase
This is relative lower precision measuring phase. The resolution of
measuring phase I of WL11, WL10, WL9 Digital Level Meter is
0.01mm/m, suit to inspect the flatness, straightness of products with
precision grade“1”.
Reminding
➜
Different resolutions of different measuring phase suit to different accuracy
inspections. Please choose proper product and measuring phase accordingly.
➜
The higher measuring phase limit the temperature and environment strictly.
If environment and temperature could not be ensured as requested, the apparatus may
be not stabilized.
Zero Setting
Being at this mode, press button
and
, the Level Meter
zero position could be set.
(1) At the working surface being inspected, set Zero at any position (relative zero position).
(2) Absolute zero setting
(3) Nice zero setting could be available by pressing the number descending or ascending.
Reminding
➜
Regarding to the detailed process of zero setting for Digital Level Meter, please read
Segment “Apparatus Adjustment” on page 16.
Voltage
Function of measuring the power voltage automatically. LED
displayer shows the voltage (V) of the batteries being in use. When
the voltage is lower than 4V, low voltage warning
will flash.
10
Status
Pro
–
µLevel User Guide
GETTING STARTED