16 PKG08762-UM3754
ENGLISH
SETTING LIMITS (GO/NO GO FUNCTION)
1. Select the units to be displayed.
2. Press the
SHIFT/SET
button. The SET
icon will appear in the display Fig. 8.
3. Press the
LIMITS
button. The MIN LIMIT
icon will appear on the display Fig. 9.
4. Press the
SHIFT/SET
button. The LIMIT
icon will flash.
5. Adjust the Height Gage to the desired minimum value.
6. Press the
SHIFT/SET
button to capture
the minimum value. The LIMIT icon will
stop flashing.
7. Press the
LIMITS
button again. The MAX
LIMIT icon will appear in the top middle of
the display Fig.10
8. Press the
SHIFT/SET
button. The LIMIT
icon will flash.
9. Adjust the Height Gage to the desired
maximum value.
10. Press the
SHIFT/SET
button to capture
the maximum value. The LIMIT icon will
stop flashing.
11. Press the
LIMITS
button. The LIMIT and
the SET icon will remain on the display.
The MAX and MIN icons will stay off as long as the reading is within the set limits.
If the measurement is beyond the set limits the display will flash and either the
MIN or MAX LIMIT icon will indicate the direction the limit has been exceeded.
12. To exit Press the
LIMITS
button.
FIG. 8
FIG. 9
FIG. 10
Содержание 3754