Measurement units:
metric and Imperial
2)
Receiving gain:
LOW and HIGH
1.
The LOW is mainly used for measuring coarse material
with high scatter and small sound absorption, such as cast
aluminum, cast copper and other metallic parts.
3)
Display resolution:
0.1mm(LOW) and 0.01mm(HIGH)
4)
Minimum capture measurement
OFF and ON
1.
To
capture
min.
measurement is to trap the
minimum value in a group of
measured values. When the
probe couples with the work
piece, it will display actual
measurement,
when
the
probe is taken away, it will
display the min. value of
measurement carried out a moment ago, and the MIN indicator
for the minimum value will flash several seconds. If you continue
the measurement when the MIN is flashing, the former
measurements will continue to take part in the min. value
capturing. If you carry out measurement after MIN indicator stops
flashing, the min. value capturing will begin from then on.
2.
When the Min. capture function is ON, LCD will have MIN
indication
5)
2-Point Calibration:
OFF and ON
1.
Select two standard samples of the same material with
workpiece to be measured, among which, one has a thickness
equal to or slightly higher than the tested piece, and the thickness
of another test piece is slightly lower than the tested piece.
2.
Before carrying out 2-point calibration, please turn off Min.
Capture function, perform the Erase CAL Data function in the
Memory Manager, and erase the former calibration record.
3.
Set the 2-point calibration is ON.
4.
Press
, key return main display.
5.
One can carry out 2-point calibration at any time during
measurement.