Operation   |   7 

6   |   Operation

Charging Your Hearing Aids

•  Place your hearing aids in the charger:

 

–  Premium Standard Charger: place hearing aids 

into charging ports with earbuds/earmolds 
resting in the reservoir�

 

– Premium Custom Charger: place hearing aid 

faceplate facing downward so battery contacts 
on hearing aid align with charging posts within 
charger�

•  When hearing aids are placed in the charger, they 

will automatically turn off and begin to charge� 
When hearing aids are removed from the charger, 
they will automatically power on�

•  The LEDs corresponding to each hearing aid:

 

–  Glowing Green = Charging

 

–  Solid Green = Fully Charged *

 

–  Blinking Red = Fault State� To reset, remove 

the hearing aids from charger, wait until LED 

turns off and re-insert aids� If Fault State 
continues, call your hearing professional�

•  Charging occurs with the lid open or closed; 

hearing aids will be fully charged in under 3 ½ 
hours

•  It is safe to store hearing aids in the charger after 

they are fully charged and any time that you are 
not wearing them�

•  If you will not be wearing your hearing aids for 

an extended period of time (i�e�weeks), remove 
both the plug from the premium charger and the 
hearing aids from the charging ports/posts�You 
will need to manually power off the hearing aids 
by pressing the user control on each hearing aid 
for three seconds�

•  To illuminate LEDs when the premium charger 

is not plugged in, remove a hearing aid from its 
charging port/post for three seconds and then  
re-insert it� LEDs illuminate for 10 seconds before 
turning off� 

•  The desiccant puck provided with the Premium 

Standard Charger will be effective for 3–6 months 
depending on moisture in the hearing aids and 
humidity� When the desiccant puck fades from 
blue to white, it should be replaced�

* If you are charging without the cord, the LEDs will turn off when the hearing aids are 

fully charged to preserve the on-board battery.

Содержание Premium Custom Charger

Страница 1: ...O P E R AT I O N S M A N U A L Standard Charger Custom Charger Premium ...

Страница 2: ... 11 Regulatory Information Safety Information 13 FCC Information 15 The Premium Standard Charger is designed to be a portable charger for standard style lithium ion rechargeable hearing aids The Premium Custom Charger is designed to be a portable charger for custom style lithium ion rechargeable hearing aids Both premium chargers work when plugged into a wall or on the go with an on board battery ...

Страница 3: ...w 5 3 2 1 6 7 Features Controls and Identification 1 Desiccant puck holder 2 Cleaning brush 3 Earbud earmold reservoir 4 Charging ports posts 5 Hearing aid charging LEDs 6 On board battery LEDs 7 Power supply port Premium Standard Charger only 5 3 2 4 4 ...

Страница 4: ...hearing aids will be fully charged in under 3 hours It is safe to store hearing aids in the charger after they are fully charged and any time that you are not wearing them If you will not be wearing your hearing aids for an extended period of time i e weeks remove both the plug from the premium charger and the hearing aids from the charging ports posts You will need to manually power off the heari...

Страница 5: ...in 4 Solid 75 3 Solid 75 2 Solid 50 1 Solid 25 1 Blinking Charge Low Power On Off ON The hearing aids will power on automatically after removing them from the charger There is a delay that allows you time to insert your hearing aid into your ear ON If the hearing aid is manually powered off pressing the user control for three seconds will power it on The delay exists to give you time to insert the...

Страница 6: ...harging hearing aids no LEDs are illuminated On board battery save mode To illuminate LEDs remove hearing aid from its charging port post for three seconds then re insert it LEDs will remain illuminated for 10 seconds Troubleshooting Guides 11 Premium Charger Care Keep your charger clean Heat moisture and foreign materials can result in poor performance Using the provided cleaning brush keep the c...

Страница 7: ...ories 10 rH 95 rH 10 C 45 C Your charger should be stored and transported within the temperature humidity and pressure ranges of 10 C 14 F to 45 C 113 F 10 95 rH and 70 kPa 106 kPa equivalent to altitudes from 1 200 ft 380 m below sea level to 10 000 ft 3 000 m The charging temperature range is between 10 C 50 F and 40 C 104 F and between 10 95 RH and 70 kPa 106 kPa Caution If the charger is hot d...

Страница 8: ... passed the following emissions and immunity tests IEC 60601 1 2 radiated and conducted emissions requirements for a Group 1 Class B device as stated in CISPR 11 Harmonic distortion and voltage fluctuations affecting the power input source as stated in Table 2 of IEC 60601 1 2 RF radiated immunity at a field level of 10 V m between 80 MHz and 2 7 GHz as well as higher field levels from communicati...

Страница 9: ...bol Meaning Applicable Standard Symbol Number Manufacturer symbol To identify the manufacturer of a product Manufacturer BS EN ISO 15223 1 2021 5 1 1 EEA EU EC European Authorised Representative symbol Authorized representative in the European Community BS EN ISO 15223 1 2021 5 1 2 Date of manufacture symbol To indicate the date on which a product was manufactured Date of Manufacture BS EN ISO 152...

Страница 10: ...Notes Notes ...

Страница 11: ... 2022 Starkey Laboratories Inc All Rights Reserved 86619 007 4 22 BKLT3134 01 EE SC The Starkey logo and Starkey Hearing Technologies are trademarks of Starkey Laboratories Inc ...

Отзывы: