3-36
Scan - Rel
SR844 RF Lock-In Amplifier
number of interpolated points is N – 2 .
Completing the scan parameter setup procedure does
not
start a scan. The frequency
reference remains unchanged (Internal or External reference mode) after all scan
parameters have been set.
Start/Step
This key is used to step through the scan frequencies. If no scan has been setup, the
factory preset values are used, namely Start = 100 kHz, Stop = 100 MHz, N = 4, which
gives interpolated frequencies of 1 MHz and 10 MHz.
The SR844
must
be in Internal Reference mode to perform a scan.
Start/Step
has no
effect unless the unit is in Internal Reference mode.
Successive presses of
Start/Step
perform the following functions.
Keypress
Description
1st
If a scan is not in progress, pressing the
Start/Step
key will begin one, if
the SR844 is in Internal reference mode. The frequency is set to the Start
frequency, F
start
, and the
SCAN MODE
indicator is turned on.
2nd ... Nth
Subsequent keypresses step to the next scan frequency, F
i
. When the
frequency reaches the Stop frequency, F
stop
,the
SCAN DONE
indicator
turns on.
N+1st
Another keypress exits the Scan Mode while remaining at the Stop
frequency. The
SCAN MODE
and
SCAN DONE
indicators both turn
off.
N+2nd
The next keypress begins another scan, just like the 1st keypress above.
Important!
During a frequency scan, changing some front panel parameters will exit Scan Mode.
These are (1) changing the internal frequency and (2) switching to External Reference
mode.
OFF
[Shift–
Scan Set]
This key sequence terminates Scan Mode. The
SCAN MODE
and
SCAN DONE
indicators are turned off. It has no effect if
SCAN MODE
is off.
Storing and Using Rel Values
This section discusses the storage of Rel Values at the scan frequencies. The Scan and Rel features work
together in a straightforward manner. First, use the Scan keys to specify the frequency scan (Start, Stop
and N) as described previously. Then, use
Start/Step
to step through the scan frequencies while using the
Store keys to store Rel Configurations and Values at each frequency. Changing scan parameters clears all
stored Rels.
XY Rel Values are actually stored X and Y offsets. R
θ
Rel Values are stored offsets for R[dBm] and
stored settings of the reference phase. Rel Values are stored along with the current measurement
configuration. This is because offsets are generally valid only for a specific configuration and are not
Содержание SR844
Страница 10: ...viii SR844 RF Lock In Amplifier...
Страница 12: ...1 2 Getting Started SR844 RF Lock In Amplifier...
Страница 32: ...2 2 SR844 Basics SR844 RF Lock In Amplifier...
Страница 60: ...3 2 Operation SR844 RF Lock In Amplifier...
Страница 102: ...3 44 Shift Functions SR844 RF Lock In Amplifier...
Страница 108: ...4 6 Index of Commands SR844 RF Lock In Amplifier...
Страница 144: ...4 42 Example Program SR844 RF Lock In Amplifier...
Страница 146: ...5 2 Performance Tests SR844 RF Lock In Amplifier...
Страница 150: ...5 6 Performance Tests SR844 RF Lock In Amplifier...
Страница 156: ...5 12 Performance Tests SR844 RF Lock In Amplifier...
Страница 158: ...5 14 Performance Tests SR844 RF Lock In Amplifier...
Страница 162: ...5 18 Performance Tests SR844 RF Lock In Amplifier...
Страница 166: ...5 22 SR844 Test Record SR844 RF Lock In Amplifier...
Страница 168: ...6 2 Circuitry Parts Lists and Schematics SR844 RF Lock In Amplifier...
Страница 246: ...Parts Lists SR844 RF Lock In Amplifier 6 80 Schematic Diagrams...