Stanford Research Systems
PS300 Series High Voltage Power Supplies
Remote Operation
3 – 8
TMOD(?) {
i
}
Trip Mode
The
TMOD command sets (queries)
the trip reset mode {to
i
}. The
value
i
= 1 sets manual trip reset, while the value
i
= 1 sets the trip
reset mode to automatic.
Example:
TMOD 1
VLIM(?) {
x
}
Voltage Limit
The
VLIM
command sets (queries) the value of the voltage limit
{to
x
}, in volts. The sign of value
x
must
match the polarity of the
power supply. As with front-panel control, the
VLIM
value must be
greater than or equal to the
VSET
value or an execution error will
be
returned.
Example:
VLIM?
-2.0000E4
VSET(?) {
x
}
Voltage Set
The
VSET
command sets (queries) the value of the voltage se ing
{to
x
}, in volts. If SMOD 1 (rear-panel control) is enabled, then
se ing the VSET command returns an error. The value
x
must
match the polarity of the power supply.
When SMOD = 0 (front-panel control), the VSET? query returns
the set parameter (not the measured output voltage,
see VOUT?
§
3.4.4
). When SMOD = 1 (rear-panel control), the VSET? query
returns the high voltage se ing programmed by the rear-panel
HVSET input. As with front-panel control, the VSET value must
be less than or equal to the VLIM value or an execution error will
be
returned.
Example:
VSET?
19555
3.4.6 Interface Control Commands
*RST
Reset
The
*RST
common command resets the power supply to its default
con
fi
guration. It is the same as holding the [CLR] bu on
depressed at power on.
Example:
*RST
Содержание PS355
Страница 4: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Contents ii ...
Страница 10: ...Phone 408 744 9040 www thinkSRS com PS300 Series HighVoltage Power Supplies General Information viii ...
Страница 16: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Instrument Overview 1 6 ...
Страница 22: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Guide to Operation 2 6 ...
Страница 38: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Remote Operation 3 16 ...
Страница 42: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Troubleshooting 4 4 ...