Stanford Research Systems
PS300 Series High Voltage Power Supplies
Remote Operation
3 – 2
Symbol
De
fi
nition
i
Unsigned
integer
x
Assigned value
(?)
Required for queries; illegal for set commands
var
Parameter always required
{
var
}
Parameter required for set commands; illegal
for
queries
[
var
]
Optional for both set and query commands
Output Commands
HVOF
3 – 6
High Voltage O
ff
HVON
3 – 6
High Voltage On
IOUT?
3 – 6
Output Current
VOUT?
3 – 6
Output Voltage
Setting Control Commands
*RCL
i
3 – 6
Recall
*SAV
i
3 – 7
Save
ILIM(?)
{
x
}
3 – 7
Current Limit
ITRP(?)
{
x
}
3 – 7
Current Trip
SMOD(?)
{
i
}
3 – 7
V Set Mode
TCLR
3 – 7
Trip Clear
TMOD(?)
{
i
}
3 – 8
Trip Mode
VLIM(?)
{
x
}
3 – 8
Voltage Limit
VSET(?)
{
x
}
3 – 8
Voltage Set
Interface Control Commands
*RST
3 – 8
Reset
*IDN?
3 – 9
Identify
*OPC(?)
3 – 9
Operation Complete
LERR?
3 – 9
Last Error
Status Reporting Commands
*CLS
3 – 10 Clear Status
*ESE(?)
i
3 – 10 Standard Event Status Enable
*ESR?
[
i
]
3 – 10 Standard Event Status
*PSC(?)
{
i
}
3 – 11 Power-On Status Clear
*SRE(?)
{
i
}
3 – 11 Service Request Enable
*STB(?)
[
i
]
3 – 11 Status Byte
3.1 Index of Common Commands
Содержание PS355
Страница 4: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Contents ii ...
Страница 10: ...Phone 408 744 9040 www thinkSRS com PS300 Series HighVoltage Power Supplies General Information viii ...
Страница 16: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Instrument Overview 1 6 ...
Страница 22: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Guide to Operation 2 6 ...
Страница 38: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Remote Operation 3 16 ...
Страница 42: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Troubleshooting 4 4 ...