INTRODUCTION
The procedures in this section test the performance of the DS345 and com-
pare it to the specifications in the front of this manual. The first set of tests
test the basic functionality of the DS345 from the front panel. The second set
of tests actually measure the DS345's specifications. The results of each
test may be recorded on the test sheet at the end of this section.
NECESSARY EQUIPMENT
The following equipment is necessary to complete the tests. The suggested
equipment or its equivalent may be used.
Instrument
Critical Specifications
Recommended Model
Analog Oscilloscope
350 MHz Bandwidth
Tektronix 2465
Time Interval Counter
Frequency Range: 20 MHz min.
SRS SR620
Time Interval Accuracy: 1ns min
FFT Spectrum Analyzer
Frequency Range: DC to 100 kHz
SRS SR760
Amplitude Accuracy: ±0.2 dB
Distortion: < 75 dB below reference
RF Spectrum Analyzer
Frequency Range: 1 kHz to 100 MHz
Anritsu MS2601/ HP4195A
Amplitude: ±0.5 dB
Distortion and Spurious: < -70 dB
DC/AC Voltmeter
5 1/2 Digit DC accuracy
Fluke 8840A
True RMS AC to 100 kHz
Thermal Converter
Input Impedance: 50
Ω
Ballantine 1395A-3-09
Input Voltage: 3 Vrms
Frequency: DC to 30 MHz
Accuracy: ±0.05dB
10 MHz Frequency Standard
Frequency: 10 MHz ± .001 ppm
SRS FS700
Phase Noise: < -130 dBc @ 100Hz
50
Ω
Terminator
50
Ω
± 0.2 %, 1 Watt
HP 11048C
Doubly Balanced Mixer
Impedance: 50
Ω
Mini-Circuits ZAD-3SH
Frequency: 1 - 20 MHz
1 MHz Lowpass Filter
-50 dB min at f > 15 MHz
TTE, Inc. Model J85
15 kHz Lowpass Filter
11.0 k
Ω
, 0.0015
µ
F
Homemade
5-1
PERFORMANCE TESTS
Содержание DS345
Страница 2: ......
Страница 5: ...DS345 Synthesized Function Generator iii...
Страница 20: ...Introduction 2 4...
Страница 64: ...Programming Commands 3 14...
Страница 72: ...Program Examples 3 22...
Страница 78: ...Troubleshooting 4 6...
Страница 82: ...Performance Tests 5 4...
Страница 101: ...Calibration 6 10...
Страница 109: ...Arbitrary Waveform Composer 7 8...
Страница 117: ...DS345 Circuitry 8 8...