Merlin™ PSA Help Manual
5-1
5. Lead Tests
When the Lead Tests window appears, the Merlin™ PSA begins sensing and measuring the intrinsic
signal amplitude and slew rate.
From the Lead Tests window you can:
•
.
• Reset the
,
.
• View and Store
Intrinsic Signal Amplitude and Slew Rate Measurements
.
• View and Store
• Open the
window.
.
• Perform a
Phrenic Nerve Stimulation Check
.
•
.
• In CRT devices, select the connected
• Print a report. Select the Print button. See
•
.
• Return to the device session with the Programmer button. See
Suspend the Merlin™ PSA Session
Accessed From: Lead Tests button
Enable/Disable Pacing
Select the
toggle button to turn pacing on or off for the corresponding channel. When you
enable or disable pacing for a particular channel with the Pacing button, the Merlin PSA automatically
changes the
settings, depending on the pacing status of other channels (see
and
Pacing, Mode, and V. Pacing Parameter Autoprogramming
different setting for the
You must enable pacing to measure lead impedance and capture threshold.
To enable pacing with the Pacing button, the Mode must be set to ODO, VVI, AAI, or DDD. The Pacing
button does not operate in AOO, VOO, VDD, DOO, or DDI modes.
See also:
Note
. Stimulation applied during the PSA session is ONLY
intended for use during short-term lead evaluation tests and IS NOT
intended for long term life-sustaining pacing support.
Note
When noise, defined as activity sensed by the Merlin PSA in a
refractory period, is detected, the Merlin PSA switches to
asynchronous pacing at the programmed Base Rate in that channel
until noise disappears.