7-2
Parameters
ratio, which may enhance sensing performance and may reduce oversensing T waves. Turning the Low
Frequency Attenuation filter On or Off changes the device’s sensing performance and may impact the
reported R wave amplitude. For more information on how turning the Low Frequency Attenuation
filter On or Off affects the sensing ability of the device, refer to the device manual.
For best results, the Low Frequency Attenuation setting on the PSA should match that of the selected
device.
The default setting for the Low Frequency Attenuation filter is On for all tachycardia devices except for
CRT-D devices. The default can be changed by selecting the Save as Default button on the Advanced
Settings screen.
See also:
•
Parameter Settings and Availability
Max Track Rate
The Max Track Rate parameter determines the maximum ventricular stimulation rate allowed in atrial
tracking modes.
See also:
•
Parameter Settings and Availability
Mode
The Mode parameter determines the stimulation and sensing method. The selected mode is in effect at
all times except during
. Details of each mode can be found in
See also:
•
Parameter Settings and Availability
Pacing
The Pacing parameter enables stimulation for the corresponding channel. For the Merlin™ PSA, this
parameter is linked to the Mode and V. Pacing parameters. See
Pacing, Mode, and V. Pacing Parameter
.
See also:
•
Parameter Settings and Availability
Pulse Amplitude
The Pulse Amplitude parameter determines the voltage applied to the myocardium during the
stimulation pulse.
Note
Pacing
. Stimulation applied during the PSA session is ONLY
intended for use during short-term lead evaluation tests and IS NOT
intended for long term life-sustaining pacing support.