Merlin™ PSA Help Manual
7-3
See also:
•
Parameter Settings and Availability
Pulse Width
The Pulse Width parameter determines the duration of the pacing pulse applied to the myocardium.
See also:
•
Parameter Settings and Availability
Refractory
The Refractory parameter determines the amount of time that the device is unresponsive to signals
following a sensed or paced event for the corresponding channel. The refractory period for the
corresponding channel is initiated by a sensed or paced event on that channel and either ends when
the period times out or is reset when another sensed or paced event occurs on that channel.
In AAI
, the Refractory parameter determines the atrial refractory period. In dual-chamber modes,
the Refractory parameter determines the post-ventricular atrial refractory period.
See also:
•
Parameter Settings and Availability
S1S1
The S1S1 parameter determines the interval at which the Burst Pacing stimulation pulse is applied.
See also:
•
Parameter Settings and Availability
Sensitivity
The Sensitivity parameter determines the threshold above which the device responds to intrinsic cardiac
events.
See also:
•
Parameter Settings and Availability
V. Pacing
The Ventricular Pacing parameter determines which ventricles receive stimulation and the order in
which the stimulation is applied. When you select the LV—>RV or RV—>LV settings, you can also set
the
parameter.
See also:
•