Ampere™ Generator Instructions for Use
International Edition Rev. A 100081509-en
Customizable Information Displays
Various additional options are also available to the user from the menu, including statistics on power, temperature, and impedance
from previous ablation applications, display of additional information during ablation applications, and language, audio volume, and
screen brightness control.
Safety Features
The Ampere™ Generator has many safety features and includes a separate standby button on the front panel to disable RF delivery;
automatic RF shutoff if the impedance < 50 ohms or > 300 ohms or greater than the user set impedance limit; and automatic RF shutoff
if the catheter temperature > 80 °C or if the catheter temperature exceeds the user set temperature by more than 5 °C for more than 3
seconds.
System Hardware
Components
■
Ampere™ Generator
■
Footswitch
Cables
■
Country-specific power cords
■
Equipotential Cable
■
4 Lead EGM Cable
■
Serial DB9 Cable
■
1641 Cable - Catheter Extension Cable
■
Disposable Indifferent Patch (DIP) (x2)
■
GenConnect Cable
Accessories
This accessory is available for use with the Ampere™ Generator:
■
Remote Control with power cords, equipotential cable, and fiber optic connecting cables