Ampere™ Generator Instructions for Use

International Edition Rev. A 100081509-en

System Description

Overview

The Ampere™ Generator generates radiofrequency (RF) current, at 485 KHz, to be used during RF catheter ablation procedures of 

the heart. The Ampere™ Generator operates in conjunction with an external Disposable Indifferent Patch (DIP) electrode (such as the 

1149C-LP by 3M™ or equivalent) and a compatible ablation catheter with an associated cable. The Ampere™ Generator delivers RF 

power in a unipolar mode between the ablation catheter's distal electrode and the DIP electrode. For more information regarding the 

ablation catheter, refer to its instruction manual.

The Ampere™ Generator features a color LCD screen and easy-to-use controls for setting the desired ablation parameters and for 

monitoring ablation progress. The generator can be controlled from the front panel of the main unit or by an optional remote control 

unit connected via fiber optic cables. Additional accessories to the generator include an optional footswitch the operator can use to 

turn on or off RF delivery. When connected with a compatible St. Jude Medical Cool Point™ irrigation pump, the Ampere™ Generator 

provides additional pump control options for use with an irrigated ablation catheter. The Ampere™ Generator is designed to be used 

with ablation catheters having integrated temperature sensors, and can measure temperature from a thermistor or up to 4 

thermocouples. The Ampere™ Generator provides for connection to an electrophysiologic recording system and the EnSite™ 

Velocity™ cardiac mapping system.

User Features

The Ampere™ Generator continuously and automatically monitors electrical impedance between the ablation catheter and DIP 

electrode at 485 KHz and displays the impedance on the LCD screen. Likewise, temperature from the catheter's temperature sensor is 

also continuously and automatically monitored and displayed. RF parameters of power output, temperature, duration, and an 

impedance limit are controlled from the front panel using up/down buttons. Frequently-used RF parameters can be programmed into 

one of 4 presets also accessible from the front panel. Additional display and ablation control options are accessed through a menu and 

turn knob control. RF energy can be turned on or off from a single push-button. A standby button is also available for emergency cutoff 

of RF energy. Displays and controls on the main unit are identical on the optional remote control such that changes made to settings 

on either the main unit or the remote control are reflected on both.

TempGuard™ Feature

The Ampere™ Generator features a user selectable temperature control algorithm, referred to as the TempGuard™ feature. When the 

TempGuard™ feature is selected (through the Ablation Parameters menu), the generator operates in temperature control mode, and 

power output from the generator is automatically modulated so that the temperature measured from the catheter remains at or below 

the user set temperature. In temperature control mode, additional options are available to control the initial response time of the 

TempGuard™ feature. Note that if insufficient power is selected while in temperature control mode, the user set temperature may not 

be achieved.

When the TempGuard™ feature is turned off, the generator operates in power control mode. In this mode, the user set power is output 

by the generator within a user settable ramp-up time, and then maintains that power. In power control mode, a user set cutoff 

temperature can be applied such that RF will automatically terminate if the catheter temperature exceeds the user set temperature.

Impedance Control

An impedance limit on the main display can be set using the up/down arrows, where RF will automatically stop if the measured 

impedance exceeds the limit. Additionally, an optional delta impedance cutoff can be selected. This delta impedance cutoff will 

automatically stop RF delivery if impedance changes by a user selected amount over a time window of 1 to 10 seconds.

Cool Point™ Irrigation Pump Integration

When used with an irrigated catheter and a Cool Point™ Irrigation Pump, the Ampere™ Generator automatically communicates with 

the pump and allows changing of pump parameters from the generator (through the Pump Parameters menu). The Ampere™ Generator 

can also be set to automatically change the irrigation rate during ablation if either the power or temperature reach a user defined level.

Содержание Ampere Generator

Страница 1: ...ul MN 55117 USA St Jude Medical Coordination Center BVBA The Corporate Village Da Vincilaan 11 Box F1 1935 Zaventem Belgium afcustomerservice sjm com www sjm com WARNING The use of this device in conjunction with a radio frequency ablation catheter as a part of the diagnosis and treatment of cardiac arrhythmias may pose an increased risk of adverse events such as cardiac perforation myocardial inf...

Страница 2: ...tch AC Power Power Plug Equipotential jack Non sterile Do not reuse On carton Transport storage temperature limitation Operating temperature Humidity limitation Weight Fragile Keep dry Manufacturer Date of Manufacture Use by date Lot number Catalog number Serial number Finished Goods number Made in U S A Not for direct patient contact Consult the Instruction for Use Dispose of hardware in accordan...

Страница 3: ...Mark Brazil Intertek Safety Agency Certification Mark On GenConnect GenConnect cable connector On EnSite Velocity Cardiac Mapping System Amplifier Fiber optic cable connector Dispersive patch cable connector Ablation catheter cable connector Telephone Facsimile Symbol Description EC REP Ablate Menu Preset Up Button Down Button Standby Button Electrogram Cable Connection Footswitch Connection USB C...

Страница 4: ...on Rev A 100081509 en Rated Current Footswitch Generator Non ionizing electromagnetic radiation Cables Cords Package Contents IP68 Footswitch enclosure is dust tight and protected against submersion in water Neutral Electrode Floating Ground Symbol Description ...

Страница 5: ... the Cool Point Irrigation pump Other important information has also been supplied for the user s convenience Precaution DO NOT attempt to operate the Ampere Generator before thoroughly reading this Instructions for Use The Ampere Generator Instructions for Use should be read understood and followed carefully For future reference keep this manual in a convenient readily accessible place Indication...

Страница 6: ...lso accessible from the front panel Additional display and ablation control options are accessed through a menu and turn knob control RF energy can be turned on or off from a single push button A standby button is also available for emergency cutoff of RF energy Displays and controls on the main unit are identical on the optional remote control such that changes made to settings on either the main...

Страница 7: ...on the front panel to disable RF delivery automatic RF shutoff if the impedance 50 ohms or 300 ohms or greater than the user set impedance limit and automatic RF shutoff if the catheter temperature 80 C or if the catheter temperature exceeds the user set temperature by more than 5 C for more than 3 seconds System Hardware Components Ampere Generator Footswitch Cables Country specific power cords E...

Страница 8: ...the ablation procedure c exercise extreme caution during ablation when in close proximity to atrial or ventricular permanent pacing leads and d perform complete pacing system analysis on all patients after ablation WARNING Ablation within the coronary arterial vasculature has been associated with myocardial infarction and death WARNING Patients undergoing left sided ablation procedures should be c...

Страница 9: ... otherwise the default values will be used CAUTION Choose the lowest possible output power for the intended purpose CAUTION Read and follow the DIP electrode manufacturer s instructions for use CAUTION Standard grounding procedures should be followed if electrosurgical instruments are used CAUTION The Ampere Generator is capable of delivering significant electrical power Patient or operator injury...

Страница 10: ...ther device s are connected 4 Consult the manufacturer for help CAUTION Follow the system installation to achieve optimal use CAUTION When used in conjunction with Cool Point Irrigation Pump 1 Refer to the Cool Point Irrigation Pump operator manual to troubleshoot the Cool Point Irrigation Pump 2 The Cool Point Irrigation Pump tubing set is required for use 3 If the communication cable used betwee...

Страница 11: ...enu 4 assignable presets and default settings 5 Standby Button Red light is on while in standby Standby message displays on screen Press to toggle On Off Standby Both the main and Remote unit Standby buttons must be Off to disengage Standby 6 EGM Connector For connecting to an EP recording system to display the intracardiac signal 7 Indifferent Electrode Connectors 2 For connecting the Ampere Gene...

Страница 12: ...er Watts of power to be delivered to the catheter location Arrows increase decrease power threshold in 1 Watt increments 4 User Set Temperature Temperature to be reached at the catheter sensor location Arrows increase decrease temperature threshold in 1 C increments 5 User Set Impedance Impedance between the ablation electrode and the DIP electrode s If impedance goes above set limit RF energy del...

Страница 13: ...F energy delivery power is 50 Watts only one Disposable Indifferent Patch is required Yellow Programmed RF energy delivery power is 50 Watts two Disposable Indifferent Patchs are required Catheter Gray Catheter is not connected Green Catheter is connected Remote Control Gray Remote control is not connected Green Remote control is connected and communicating with the Ampere Generator Footswitch Gra...

Страница 14: ...ord plug in 5 Fiber Optic Connector Single Fiber Optic port to connect to the optional Remote Control 6 Dual Fiber Optic Connector Dual fiber optic port to connect to the optional Remote Control 7 Fiber Optic Transceiver Connector Fiber optic port for future use 8 USB Port USB memory stick connection 9 DB 15 Serial Port female Connection for Cool Point Irrigation Pump interface 10 DB9 Serial Port ...

Страница 15: ...nit or Remote Control enters standby mode which disables all output until the button is released Both the main unit s and the Remote Control s Standby button must be released to terminate standby mode Ablation Continues from 1 999 seconds Safety Class Class I Type CF according to IEC 60601 1 Operating Frequency 485 kHz Operating Parameters Values are digitally displayed on the Ampere Generator fro...

Страница 16: ...erature 10oC to 35oC Relative humidity 20 to 90 non condensing The operating ambient altitude range shall be 0m to 3000m 9843 The Ampere Generator is intended for use in the electromagnetic environment specified below The customer or the user of the Ampere Generator should assure that it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF emissions CISP...

Страница 17: ...EN61000 4 5 IEC 1000 4 5 1 kV differential mode 2 kV common mode 1 kV 2 kV Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 95 dip in VNOM for...

Страница 18: ...by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects...

Страница 19: ...pment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 117 117 233 10 369 369 737 1 1 167 1 167 2 33 10 3 69 3 69 7 37 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the e...

Отзывы: