Connection Icons
The Connection Icons indicate the state of connected accessories:
Figure 3. Connection Icons
Icon
Name
State
Description
TempGuard™
Feature
Gray
• The TempGuard™ feature is not enabled.
White
• The TempGuard™ feature is enabled and ready.
Green
• The TempGuard™ feature is active.
DIP X2
Gray
• Programmed RF energy delivery power is
≤
50 Watts — only one Disposable
Indifferent Patch is required.
Yellow
• Programmed RF energy delivery power is > 50 Watts — two Disposable
Indifferent Patchs are required.
Catheter
Gray
• Catheter is not connected.
Green
• Catheter is connected.
Remote
Control
Gray
• Remote control is not connected.
Green
• Remote control is connected and communicating with the Ampere™
Generator.
Footswitch
Gray
• Footswitch is not connected.
Green
• Footswitch is connected and communicating with the Ampere™ Generator.
Pump
Gray
• Pump is not connected.
White
• Pump is connected and ready.
Green
• Pump is running.