10
Maintenance
Trouble Shooting
The digital surface Resistance Meter requires little maintenance. There are no user servieable parts. If the
meter requires service beyond cleaning the electrodes or replacing the batteried, please contact SHISHIDO
Local agents or SHISHIDO Japan.
Battery replacement
Replace the batteires onece the battery life indicator is empty. Open the battery cover at the back side to
replace the batteries. The meter uses four AA alkaline batteries. Ensure that the batteries polarities are
oriented in the correct direction to avoid any possible circuit damage.
Cleaning the Digital Surface Resistance Meter
The area surrounding the test jacks at the tp end of the meter shoud be wiped with aclean, isopropanol-
alcohol and affect the meter's accuracy at high resistance. The frequency of cleaning this area at least
onece a month. Cable jackets should also be cleaned in this fashion.
In case the device does not operate correctly, it may be the result of one of the following
[The display pnel is no any indication when the power is turned on.]
Check the battery installation first. No battery, correct polarity, battery life.
[Measurement not available]
Confirm the cable connection to the probe.
[Others]
Even you checked above items, stil have trouble, please try to contact our local agent or
Shshido Japan.
Specifications
Measuring range
10 Volt supply: 1×10
3
~×10
6
Ω
100 Volt supply: 1×10
6
~1×10
12
Ω
Accuracy
±10
%
(Over 10
11
Ω ± 20%)
Voltage supply
10 Volts, 100 Volts (±5%)
Memory capacity
100 data
Power supply
4 pcs AA Alkaline batteries
Battery Life
Approximately 1500 measurements
Weight
400
g
Dimensions
100mm × 210mm × 32mm
Operating Environment
Temperature: 5 ~ 40°C Huidity: Under 80%RH