SR Instruments SRScales SRV714 Series Скачать руководство пользователя страница 12

Model SRV714 Series 
Operating and Service Manual 
Part No. MAN714_200601 

Page 12 of 14 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

email:  

[email protected]

     

  

Copyright 2020

 

S

Instruments, Inc. 

Low Limit

Applied 

Load

High 

Limit

0

0

0

990

1000

1010

2970

3000

3030

5940

6000

6060

8910

9000

9090

14850

15000

15150

Calibration Chart (+/- 1%)

CALIBRATION (CON’T) 

STEP 10:

  Ensure nothing is in contact with the platform then press 

the “

S1

” button when the display reads “

ZERO

”. 

STEP 11:

  When the display reads “

FULL

”, place the weight from 

Step 1 on the platform and press the “

S1

” button. 

STEP 12:

  Press the “

S1

” button once more to save. To exit without 

saving, press “S4” to select “

QUIT

”, then press the “

S1

” 

button. 

STEP 13

:  When  finished  making  adjustments,  remove  all  weight 

from the scale.

 

STEP 14:

  Remove the center Top Plate, disconnect the two  cables 

to the Transducers. Re-attach the Display to the platform 
and securely tighten screws. Re-attach plastic cover to Display. 

NOTE:

 Tighten screws only until rubber seal starts to compress. 

STEP 15:

  Connect the two Display cables to Transducers. Position Top Panel and slide back in place 

over the Transducer assembly pins. 

STEP 16:

  Press  the  “

S4” 

button  to  zero  the  platform.  When  the  display  indicates  “

0.0

”,  place  the 

calibrated weight on the scale again and check against the weight displayed (Figure 9). If the 
weight  is  not  correct,  recalibrate.  (If  calibration  cannot  be  accomplished,  call  the  Service 
Department.). 

 
 
 
 
 
 
 
 
 
 
 

TROUBLESHOOTING 

 

SYMPTOM 

REASON/CORRECTIVE ACTION 

Weight reading is much lower than 
expected. 

Check that the platform is clean underneath, stand on 
each  of  the  four  corners  to  see  if  one  corner  is  not 
weighing correctly. 

For additional information or assistance, telephone our Service Hotline: 1-800-654-6360 

or e-mail: 

[email protected]

 

 

CAUTION 

The integrated circuits and semiconductors 

on the printed circuit boards may be 

damaged by electrostatic discharge (ESD). Be 
sure to use proper handling precautions at all 

times. 

Figure 9: Calibration Chart 

Содержание SRScales SRV714 Series

Страница 1: ...age 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instruments I...

Страница 2: ...ARTS and ACCESSORIES 6 SYSTEM DESCRIPTION and INTENDED USE 6 MAINTENANCE and CLEANING 7 STORAGE and TRANSPORTATION 7 THEORY OF OPERATION 7 SPECIFICATIONS 8 BASIC SYSTEM OPERATION 9 BATTERY REPLACEMENT...

Страница 3: ...ruments com Copyright 2020 SInstruments Inc PACKING CHECKLIST Model SRV714 Large Animal Platform Scale DESCRIPTION QUANTITY TOP PANELS 47 in 119 cm x 31 75 in 81 cm 3 ea TRANSDUCERS 95 3 8 in 242 cm x...

Страница 4: ...to battery holder imprinted instructions Attach the battery holder cover and tighten the thumb screws Re secure the plastic cover with the four 4 screws STEP 3 Remove the other two end panels from th...

Страница 5: ...les are in line with the connectors on the transducer tubes Figure 4 STEP 8 After the transducers are plugged in carefully align the center panel with the four 4 alignment pins that are located in the...

Страница 6: ...Scale derives its power from six 6 each D cell batteries Weight is displayed with a displayed resolution of 5 pounds or 2 kilograms The SRV714 Large Animal Platform Scale is equipped with a Bluetooth...

Страница 7: ...to a digital signal by an analog to digital converter The digital signal is transferred to a micro controller where it is filtered converted to appropriate units and displayed on a liquid crystal dis...

Страница 8: ...BLUETOOTH ON TIME 10 minutes from last button press AVERAGING Automatic digital filter POWER SUPPLY Six 6 D cell batteries CALIBRATION Calibration is traceable to NIST standards OPERATING CONDITIONS N...

Страница 9: ...ots in the upper right corner of the SR App and select to Connect a device Figure 6B STEP 5 Select the SRV714 Bluetooth module from list of devices Figure 6C If the device is not listed the display ma...

Страница 10: ...ry replacement is needed an indicator will appear on the display STEP 1 Remove the center Top Panel Figure 4 and disconnect the two 2 cables from the Transducers Turn the Top Panel over and remove the...

Страница 11: ...the Display housing is out from under the scale assembly STEP 5 Press the calibration button S2 Figure 8 The display will read CAL STEP 6 Press button S4 to scroll through menu options until FULL is d...

Страница 12: ...Display to the platform and securely tighten screws Re attach plastic cover to Display NOTE Tighten screws only until rubber seal starts to compress STEP 15 Connect the two Display cables to Transduc...

Страница 13: ...on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event sha...

Страница 14: ...AN714_200601 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S By...

Отзывы: