Model SR2555 Stand-On Scale
Rev 991222
0.0
W
SECTION II
S
S
2.1
Special Features
CONTINUOUS WEIGH
and
AUTO-HOLD
The SR2555 has the capability of displaying patient weight in two different ways.
2.1.01
CONTINUOUS WEIGH -
When the SR2555 is shipped it comes in the Continuous Weigh Mode. This means
that when the patient is being weighed, the weighing surface remains
ACTIVE
and the display may show
variations depending on the movement of the patient. To
LOCK
the displayed reading you must press
the
"MEMORY/RECALL"
button. This will also store the weight data in memory if you wish to recall it.
2.1.02
AUTO-HOLD -
The Auto-Hold Mode will automatically
LOCK
the displayed weight data and store it in
memory as the last weight. The Auto-Hold Mode is activated as follows:
Press and Hold the
"MEMORY/RECALL"
button for approximately 5 sec. until the display indicates
"AUTO-HOLD ENABLED"
then release the button.
To return to the Continuous Weigh Mode simply repeat the same procedure. The display will indicate
"CONTINUOUS WEIGH ENABLED".
2.1.03
Procedure for Weighing Patients in AUTO-HOLD mode
1. Press the
"0"
button to set the zero.
2. Position the patient on to the scale.
3. When the patient's motion has stabilized, press the
"WEIGH"
* button.
* Note: In the Auto-Hold mode you
MUST
press the "WEIGH" button to display the weight.
If the display turns off before you view the weigh data or if you wish to re-weigh the patient simply press the
"WEIGH"
button once again while the patient is still on the scale.
IMPORTANT:
If after the display turns off you wish to recall the last stored weight you
MUST
press the
"MEMORY/RECALL"
button
FIRST
. Otherwise the last stored weight will be reset.
Page 5