SR Instruments SR Scales SR2555 Скачать руководство пользователя страница 5

Model SR2555 Stand-On Scale

Rev 991222

1.4

ASSEMBLY

IMPORTANT: 

All of the scale systems

are calibrated at our factory as a

complete system. Components should

not be interchanged with other

systems. To assure optimal accuracy

and reliability, make sure the serial

numbers on the readout (2) and the

base (1) match.

1.

 

Insert mast(5) into base assembly

(1) as shown.

2.

 

Remove the 4 set screws (10) from

the base collar and apply loctite

®

to all set screws

3.

 

Make sure the mast is seated firmly

in the base assembly. Re-install the

set screws and tighten.

4.

 

Slide the cable from the display

assembly (2) down the mast (5)

and out through the base.

5.

 

Insert the mast (5) into the collar of

the display assembly (2) and seat

firmly. Turn the readout assembly

(2) so that it is aligned with the

base assembly.

6.

 

Apply loctite

® 

to all set screws and

tighten display assembly in place.

7.

 

Plug the display assembly cable

into the base assembly as shown

and secure the cable in place with

the strain relief clamp provided at

the bottom of the collar in the

base assembly. Make sure the

excess cable is fed back into the

mast (5) before tightening the

strain relief clamp.

8.

 

Remove the battery cover (8) and

install the 6 “LR14” (C cell) batteries

provided. Replace the cover.

OPTIONAL HANDRAILS –

Install the handrails with the hardware

provided. Be sure to use the lock

washers (12) and apply loctite

® 

to all

screws.

1.

 

Attach the LEFT and RIGHT handrails (3,4) to the bottom of the readout assembly (2) first and then to the base (1)

as shown.

OPTIONAL HEIGHT BAR –

1.

 

Insert the Height Bar (6) from the top making sure it is seated firmly at the base.

2.

 

Align the top extension perpendicular to the readout

3.

 

Apply loctite

® 

to all set screws and tighten in place

Page  3

Содержание SR Scales SR2555

Страница 1: ...91222 PART No MAN2555 S by SR Instruments Inc Model SR2555 Stand On Scale Operating Manual SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 1 716 693 5977 Fax 1 716 693 5854 URL www srscales...

Страница 2: ...ation 1 3 Assembly 1 4 SECTION II Button Functions Basic System Operation 2 0 Special Features Continuous Weigh Auto Hold 2 1 SECTION III Calibration 3 0 Periodic Calibration Check 3 1 ESD 3 2 Calibra...

Страница 3: ...tomatically power down in 30 seconds to conserve battery life 1 2 09 Averaging An Automatic Digital Filter samples the weight data to filter motion 1 2 10 Power Supply Six LR 14 C cell Alkaline Batter...

Страница 4: ...lope analog to digital converter Analog board U5 The converter integrates the analog signal onto the integrating capacitor Analog board C9 over a short interval The integrating capacitor is then disch...

Страница 5: ...d with the base assembly 6 Apply loctite to all set screws and tighten display assembly in place 7 Plug the display assembly cable into the base assembly as shown and secure the cable in place with th...

Страница 6: ...able to view the weight data simply press the WEIGH button to wake the display and view the patient s weight The display will indicate CURRENT WEIGHT XXX X POUNDS or KILOGRAMS 2 0 4 MEMORY RECALL Pre...

Страница 7: ...d Mode is activated as follows Press and Hold the MEMORY RECALL button for approximately 5 sec until the display indicates AUTO HOLD ENABLED then release the button To return to the Continuous Weigh M...

Страница 8: ...within 0 1 of the calibrated weight plus or minus 1 digit of reading 6 Remove the weight and repeat steps 3 5 to be sure of consistent results 7 Press the SAVE button to save the calibration data The...

Страница 9: ...Model SR2555 Stand On Scale Rev 991222 SECTION III S S 3 6 SCHEMATIC DIAGRAM Page 7...

Страница 10: ...U 6 E X S E 3 3 3 1 C R Y S T A L9 6 0 0 0 K H z D IG I K E Y 1 X 1 E D 1 N 5 8 1 7 D I O D E1 N 5 8 1 7 1 D 1 E S R U E 0 9 0 S W IT C H P O L Y R A Y C H E M R U E 0 9 0 1 F 1 E D 1 N 9 1 4 D I O D...

Страница 11: ...Model SR2555 Stand On Scale Rev 991222 SECTION III S S 3 8 COMPONENT LAYOUT Page 9...

Страница 12: ...Model SR2555 Stand On Scale Rev 991222 3 9 PRODUCT LABELING...

Страница 13: ...Model SR2555 Stand On Scale Rev 991222 S By SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 58 54 www srscales com email sri srinstruments com...

Отзывы: