GUIDANCE AND MANUFACTURER’S DECLARATION
ELECTROMAGNETIC EMISSIONS
The Free Action is intended for use in the electromagnetic environment specified below. The customer or the user of
the Free Action should assure that it is used in such an environment.
EMISSIONS
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
RF emissions
CISPR 11
Group 1
The Free Action uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Free Action is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Conform
Class A
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Conform
GUIDANCE AND MANUFACTURER’S DECLARATION
ELECTROMAGNETIC IMMUNITY
The Free Action is intended for use in the electromagnetic environment specified below. The customer or the user of
the Free Action should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601 TEST LEVEL
COMPLIANCE LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
and dc lines
±2 kV for power supply
lines
Mains power quality should be that of a typical
commercial or hospital environment.
±1 kV for input/output lines
Not applicable for input/
output lines
Surge
IEC 61000-4-5
±1 kV line(s) and neutral
± 2 kV line(s) to earth
±1 kV line(s) and neutral
N/A(§7.2 EN61000-4-5)
Voltage
dips, short
interruptions
and voltage
variations
on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
Mains power quality should be that of a typical
commercial or hospital environment. If a dips or an
interruption of mains power occurs, the current of
the Free Action may be dropped off from normal
level, it may be necessary to use uninterruptible
power supply or a battery.
40 % U
T
(60 % dip in U
T
)
for 5 cycles
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5s
<5 % U
T
(>95 % dip in U
T
)
for 5s
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE - U
T
is the a.c. mains voltage prior to application of the test level