- 16
-
- 15
-
RECOMMEND
A
TIONS OF USE
- Y
ou can restart a program’
s cycle on another muscle zone
or change the program while remaining on the same muscle zone,
with
out e
xceeding however 40 min use per muscle zone and per day
.
- For optimum use, refer to the diagrams shown in this leaflet.
- Clean the unit and its cables with a cloth dampened with a disinfecting solution (mercr
yl type) diluted in water at 1/10th.
- The adhesive electrodes have a lifetime of about 40 uses, you can order them directly from the company headquarters or from y
our reseller
SPOR
T-ELEC Institut
- BP35 - 31 Rue du V
al Breton - 27520 BOURGTHEROULDE - FRANCE
Tél. 02 32 96 50 50 - Fax. 02 32 96 50 59 - www
.sport-elec.com / [email protected]
SAFETY CHECK
Ever
y 24 months, check the following points :
1 - Check for superficial degradation of the equipment
2 -
Check proper operation of the unit :
* legible display
* switch, contact
3 -
Measure electric values :
* with a 500-ohm load, the current must not exceed 100 V
.
4 -
Check reliability of accessories :
* adhesive electrodes in good condition (no tears, no material defects)
* wire of adhesive electrodes in good condition.
IV
. Recommendation of use
W
ARNINGS :
Do not use the device by driving or by working on a machine.
Do not use this electro stimulation unit in a damp environment, a bathroom, a sauna, or close to electric machiner
y connected t
o mains
and earthed,
or even piping earthed.
The simultaneous connection of a patient to a high frequency surgical
unit may
cause burns at the points of contact of the stim
ulator’
s electrodes,
and the
stim
ulator may be damaged. Operation in close proximity of a short wave ther
apy unit may cause instability in the stimulator’
s output
power
.
PRECA
UTIONS OF USE
* Do not use the unit on your face
* Do not leave within reach of children
* Ask your doctor’
s advice :
- if you have electronic or metallic implants (pins, etc.)
- in case of skin condition (wounds,…)
The unit is recommended for physiotherapists for functional re-education.
RECOMMEND
A
TIONS :
Please apply regulations enforced in your countr
y
when disposing your product at its end of life.
Therefore, please discard it in a location planned to this effect,
ensuring it is disposed in an environment-friendly manner
•
Use only on those parts of the body for which the accessories were designed
.
•
Functions only on batteries
•
Not to connect to another source of power
ST
AND
ARDS
Unit : EN60601-1 EN60601-2-10
Hydrophilic and hypoallergenic adhesive
electrodes
(ø 50 mm) : 93/42/CEE
Classification type BF
(Directive 89/336/CEE)
L’APP
AREIL :
Type of current
Number of programmes
Number of phases variations
Number of adjustable independent outputs
Current
Frequency range
Pulse width range
Power
Unit dimensions and weight
Automatic stop at end of each program
Activation safety
Detection absence of electrodes or of contact
Indicator of power
Indicator of weak and/or defective batteries
Indicator absence contact electrodes
Indicator of the programs
Temperature of storage
Relative humidity of storage
Temperature of use
Relative humidity of use
Connection device / electrodes
Caution : Read the manuel thoroughly
RoHS : The unit are free of materials likely to have a
negative impact on the environment
Guarantee
V.
T
ECHNICAL
SPECIFICA
TIONS
UNIT :
Dual phase
4 programs
79
2
From 0 to 70 mA - From 0 to 1000
Ω
1 to 100 Hz
160 to 320 µs
3 x 1,5 V batteries type LR6 (not provided)
69 x 43 x 87 (h) mm - 106 g
Yes - about 4 min 30 sec
100%
yes detection independent on each of the 2 outputs
bar graph - LCD display
yes - LCD display
yes - LCD display
pictograms - LCD display
0°C to + 45°C
10% to 90%
+ 5°C to + 45°C
20% to 65%
wires - plugs ø 2 mm male
2 years
R
oHS
(2002/95/EC)
MDE BC4 707 19/09/07 10:42 Page 15