1300-87072-1A, REV 0

 

Spex® SamplePrep 

Page 8 of 26 

4.0 SETTING UP 

The 2050 Genomax weighs 100 pounds (45.5 kg). The lid opens from the front with the handle 

in the center. The AC Input module, fuse tray, and power cord receptacle are located on the 

rear of the Genomax cabinet, toward its bottom. The power (On/Off) button is on the front right, 

as shown in Figure 2. The Manual Lid Release Latch is located on the rear of the unit at the top 

left. There are gas springs on both sides of the cabinet to control the movement of the lid. The 

touchscreen controller is affixed to the front of the unit. Under the front edge of the controller is 

an USB port to export run history or upgrade system software (Figure 1). 

 

4.1 Electrical Connection 

The detachable power cord should be plugged firmly into its inlet, then into an electrical outlet. 

Do not plug unit into a GFCI (Ground Fault Circuit Interrupter) outlet.

 Make sure that the 

electrical outlet is easily accessible in case it becomes necessary to unplug the unit. For 115 

V/60 HZ use, a 3-prong outlet fused at 20 amps is recommended. The 230 V/50 HZ power 

cord has a standard European 2-prong plug, but modification by the user may be necessary to 

meet local electrical codes. 

 

4.2 Cabinet Set Up 

To open the cabinet, grasp the red handle and raise the lid to its full upright position. The gas 

spring will hold the lid in its open position. To close, grasp the red handle and lower the lid until 

it is fully closed.  

A safety interlock will engage the Genomax only when the lid is closed. If the RUN button is 

touched while the lid is open, the Genomax will not start. The lid stays locked when the unit is 

running which prevents the user from opening it during operation. If the locking mechanism 

fails and the lid is opened, the motor will immediately stop, and movement of the clamp 

assembly will cease. 
 

4.3 Installing the #2059 Clamp Base, #2051 Clamp and #2052 Clamp 

The 

#2051

 Multi-Purpose Clamp is used with various vials, tubes, and jars (see Genomax 

accessory manual). The 

#2051

 Clamp can also be adapted to hold anything with the same 

general dimensions (5 in long x 3 in wide x 2 to 4 in high). This includes racks that hold multiple 

individual vials and other possible configurations. The 

#2052

 Titer Plate Clamp can 

accommodate a variety of titer plates (up to six). Refer to 

Section 6.0

 Loading the Clamp. 

To prevent damage during shipping, the clamp assembly is not installed at the factory. Before 

operation, this must be installed by the user. Only use the 

#2051

 Multi-Purpose Clamp or 

#2052

 Titer Plate Clamp in this unit.  

 

 

Содержание Genomax 2050

Страница 1: ...1300 87072 1A REV 0 Spex SamplePrep Page 1 of 26 2050 Genomax Automated Tissue and Cell Lyser for Biological Applications Operations Manual...

Страница 2: ...be reproduced or transmitted in any form or by any means electronic or mechanical including photocopying and recording except as may be expressly permitted by the 1976 Copyright Act or in writing fro...

Страница 3: ...5 2 3 Editing a User 11 5 2 4 Deleting a User 11 5 3 Control Panel 12 5 3 1 Starting a Programmed Run 12 5 3 2 Stopping or Pausing a Run 13 5 4 Changing Settings 13 5 5 Saved Protocols 14 5 5 1 Recall...

Страница 4: ...eatures 22 8 2 1 Protocols in Admin Mode 22 8 3 Diagnostics 23 8 3 1 Refresh Data Button 23 8 3 2 View Log Button 23 8 3 3 Export Button 23 9 0 ERROR MESSAGES 24 9 1 Lid Error 24 9 2 Rate Error 24 10...

Страница 5: ...nning time stroke rate number of cycles pause time and the RUN button pushed When the run is complete the lid is lifted and the vials or titer plates can be removed The isolation of nucleic acids from...

Страница 6: ...e 3AG 15 amp 250 V slow blow fuse for 115 V 60 Hz model 3AG 10 amp 250 V slow blow fuse for 230 V 50 Hz model Power Cord Do not plug unit into GFCI outlet Ground Fault Circuit Interrupter 115 V 60 Hz...

Страница 7: ...Follow a logical sequence of steps as you inspect the unit see figures 1 and 2 1 Inspect the electrical input module for any visible damage 2 Inspect the cabinet and lid for any visible damage 3 Ensu...

Страница 8: ...ecessary to meet local electrical codes 4 2 Cabinet Set Up To open the cabinet grasp the red handle and raise the lid to its full upright position The gas spring will hold the lid in its open position...

Страница 9: ...x Open the Genomax lid and place the clamp base on top of the shaft Using the lock washers and bolts secure the clamp base to the shaft with the Allen wrench Figure 3 IMPORTANT ONLY THE BOLTS AND LOCK...

Страница 10: ...be found on the Control Panel positioned at the upper right of the screen Touching the Home Icon allows the user to return to the Home Screen 5 2 User List To access the User List screen touch the Use...

Страница 11: ...r case the default is upper case Touch the Down arrow to switch to lower case or touch the Up arrow to revert to upper case Touching the SPACE key advances the cursor one space Touching the BACK key d...

Страница 12: ...R top of screen counts down the time remaining for a run Figure 9 To change the settings of a selected field press the arrow buttons on the CONTROL PANEL screen The left arrow buttons decrease the num...

Страница 13: ...ternatively pressing Stop will end the program 5 4 Changing Settings To change parameters run time pause time cycles and rate touch the arrow buttons on the CONTROL PANEL screen Touching the plus symb...

Страница 14: ...swipe up or down on the touchscreen This brings up vacant protocol slots on the SAVED PROTOCOLS Screen A saved protocol retains the settings for number of cycles run time pause time between cycles an...

Страница 15: ...h the protocol name to highlight the name on the Saved Protocols screen Figure 11 Then touch the Delete button A pop up window opens confirming to delete this protocol Touch YES to delete or touch NO...

Страница 16: ...vertical threaded rod through the cutout channel and the upright at the back of the clamp 3 Turn the locking nut clockwise until finger tight Check to see whether the vials can be moved from side to s...

Страница 17: ...clamp The titer plates fit snug in the clamp recess 2 Titer plates can also be stacked in the clamp to increase throughput A maximum of 6 deep well plates can be run in the 2052 clamp stacked three h...

Страница 18: ...lockwise to the ON position 8 Then use the clamp tightening tool 2052R to further tighten the locking nut 9 Close the Genomax lid and run the unit NOTE Never run the Genomax without the 2051 Clamp or...

Страница 19: ...lb 0 9 kg the maximum recommended operating rate is 1500 rpm Sample loads less than 2 lb can be run at rates up to the maximum of 1750 rpm Dynamic loads greater than 2 lb can create a rate error When...

Страница 20: ...a matter of choice which can vary with the brand of titer plate and the necessity of sealing a fluid in each well Some manufacturers of titer plates make sealing systems many technicians have also de...

Страница 21: ...fect the safety protection provided by the unit For example do not use glass vials in the Genomax Impact of the grinding media may cause the vials to break resulting in broken glass and spilled sample...

Страница 22: ...active 8 2 Administrative Lockout Features The Administrative Lockout feature passcode protected restricts user access to specific protocols or ability to modify a protocol From the Home Screen touch...

Страница 23: ...allows the user to view the diagnostic data for the unit A pop up window opens listing the date time rate and voltage data To close window touch the View Log button a second time 8 3 3 Export Button T...

Страница 24: ...he red STOP button to discontinue operation Call our service technician at 732 623 0465 for assistance 10 0 MAINTENANCE This unit has been designed to provide trouble free operation over a long period...

Страница 25: ...0 WARRANTY Spex SamplePrep guarantees its products against defects in materials or workmanship for one 1 year from the date of original shipment Repairs replacements or parts are guaranteed for 30 day...

Страница 26: ...ansportation is the responsibility of the customer Credit for returned merchandise will be issued only after goods have been received and inspected Returned goods are subject to a 25 restocking charge...

Отзывы: