6.1 Loading the #2051 Multi-Purpose Clamp
Note: The #2059 Clamp Base must be attached to the Genomax before placing the #2051
Multi-Purpose Clamp in the unit. (Section 4.3.1)
1.
Place the sample vials (in the appropriate holder or cryo-block) in the #2051 Clamp. The
sample load must be evenly distributed at all sides of the clamp to avoid excess wear to
the Genomax. In addition, the clamp lid must have a flat surface upon which to rest. If
running less than a full holder of vials, be sure to place vials at the 4 corners to support
the clamp lid.
2.
Slide the #2051 Clamp lid onto the vertical threaded rod through the cutout channel and
the upright at the back of the clamp.
3.
Turn the locking nut clockwise until finger tight. Check to see whether the vials can be
moved from side to side in the clamp. If not, the clamp is tight and samples are secure.
If the vials can be moved, tighten the locking nut (with the tool) gradually until the vials
are secure.
4.
Insert the #2051 Clamp onto the #2059 Clamp Base and lock in place by turning
locking knob clockwise to the ON position. (To remove the clamp, turn the locking knob
counterclockwise to the OFF position and lift from the #2059 Clamp Base.)
Tip:
the #2051 Clamp can also be locked into the #2059 Clamp Base prior to placing
sample vials in the #2051 Clamp.
5.
Close the Genomax lid and run the unit.
#2059 Clamp Base
Locking Knob
#2051 Clamp Lid
Locked
Unlocked
Figure 14