6875 Freezer/Mill
Page 20 of 42
SPEX® SamplePrep LLC
DEC 2015
5.4 Saved Protocols
(Cont’d)
5.4.3 Edit Saved Protocols
To edit a saved protocol, touch the protocol to highlight the name
(BLUE)
. Then touch the EDIT
button. From the Edit Protocol screen change parameters and/or name, then touch the DONE
button.
5.4.4 Recalling a program
In the
SAVED PROTOCOLS
screen, touch the protocol to highlight the name
(BLUE)
. Then touch
the LOAD button to send protocol to the
CONTROL PANEL
screen. Review the parameters. To
RUN the protocol touch the button to initiate the protocol.
6.0 OPERATION
6.1 Adding Liquid Nitrogen Coolant
Liquid nitrogen is added to the 6875 Freezer/Mill with the lid open. The 6875 should be filled
until the LN level reaches approximately 2 inches (5.1 cm) below the lip of the tub. If the liquid
nitrogen is above this level when the lid is closed, LN can splash outside the tub.
Always close the
lid slowly to avoid splashing
.
Liquid nitrogen will most likely be available either in a large tank with a flexible steel hose, or in
smaller Dewar flasks. The 6875 Freezer/Mill typically consumes 10 to 15 liters of LN during initial
cool-down, and another 4 to 6 liters per hour, depending on use.
Figure 13
– Clear Protocol Screen