24
Note...
If this secondary value continues to fluctuate, check the buffer or probe.
(Refer to TROUBLESHOOTING page 42). If CUST is selected, the
lower middle display indicates the default, 2.00. Press
HLD
to select the
buffer range needed. Press ▲ or ▼ to adjust the lower middle display to
coincide with the main display reading.
8. When the measured pH value is stable and the Ready function is
enabled (Refer to P6.0 READY ICON page 20), “READY” will appear
on the left side of the LCD. Press to confirm.
9. - For 1-Point Calibration, proceed to Step 10.
- For 2-Point, 3-Point and Multiple-Point Calibration, Repeat Steps 2
through 8 for each buffer.
10. Press
ESC.
Note...
For 2-Point, 3-Point and Multiple-Point Calibration, do not press
ESC
until after the final buffer has been calibrated.