11
SETUP MODE
The advanced Setup Mode allows you to customize the following meter
preferences and defaults:
• Memory Transmission
• Clear Memory
• Slope and Offset (pH) or Calibration Review (Conductivity)
• Buffer Solution (pH) or Cell Constant (Conductivity)
• Temperature Setting (pH)
• Ready Function
• Temperature Units
• Real Time Clock
• Reset
To enter Setup Mode, press
SET
for more than 2 seconds while the
meter is in Normal Mode.
Note...
To exit Setup Mode without saving, press
ESC
until Normal Mode
appears. If the meter is in Setup Mode, press
ESC
twice to exit. For
Conductivity, pH, and mV default settings, refer to pages 54-55.
Memory Transmission
To transfer stored data from the meter to the computer:
1. Connect a RS232 or USB cable to the rear of the meter, then connect
the other end of the cable with the D-sub connector to the computer’s
serial port. Run the software associated with this feature.
2. Press
SET
for 2 seconds to enter setup. “TR” appears on the middle of
the LCD display and P1.0 appears under “TR.”