15
SETUP MODE
Calibration Review (Conductivity Probe)
This feature allows you to review which range has been calibrated and
the last calibration value. The program reviews the probe calibration data
of Conductivity, TDS or SALT.
Note…
If the range is not yet calibrated, the LCD will display the default value.
There are 5 total calibration ranges for Conductivity, TDS and SALT.
Range 1 to 3: Conductivity or TDS value
Range 4 to 5: Conductivity, TDS or SALT value
1. Press
MODE
to select the probe program.
2. Press
SET
for 2 seconds to enter setup.
3. Press ▲ to select CAL. “CAL” appears on the middle of the LCD and
P3.0 appears on the lower portion.
4. Press
to enter P3.1. Press
to enter P3.2, P3.3, P3.4, P3.5.
5. Press
ESC
to return to P3.0.
6. Press
ESC
to return to Normal Mode.
The default values are: