LCD DISPLAY DESCRIPTION

•  The conductivity value is 

displayed on the left side of 

the LCD.

•  The temperature value is 

displayed on the upper right 

side of the LCD.

•  The total number of records saved is displayed on 

 

the lower right side of the LCD.

CONDUCTIVITY PROBE CALIBRATION

The probe should be calibrated before first use.  

•  Always use a fresh calibration solution because 

 

contaminants in the solution will affect accuracy. 

•  Calibrate at 2/3 full range under most conditions. 

 

For example, if the measuring range is 0~1999 mS, 

 

use 1413 µS solution to calibrate.

•  You will only need to do single-point calibration.  

Calibration Procedure

1.  Remove the protective cap from the conductivity probe.
2.  Rinse the probe by inserting it into de-mineralized or 

distilled water for 30 minutes.  

3

-Press  to  confirm  the 

  calibration  or  parameter 

  setting.

  LCD  DISPLAY

1.The  pH  value  will  be  displayed  on  the 

        left  side  of  the  LCD 

2.The  temperature  value  will  be    displayed 

        in  the  upper  right  corner  of  the  LCD. 

3.The  MEMO  number  which  is  the  total 

        number  of    records  saved  in  the  meter 

        will  be  displayed  in  the  lower  right  corner 

        of  the  LCD

O

C

PH

TEMP

MEMO

ATC

Conductivity  probe

4.The  conductivity  value  will  be  displayed 

      on  the  left  side  of  the  LCD.  The  Temp. 

        value  will  be  displayed  on  the  upper 

        right  corner.

5.The  temperature  value  will  be 

        displayed  in  the  upper  right  corner 

      of  the  LCD. 

6.The  MEMO  number  which  is  the  total 

        number  of  records  saved  in  the  meter 

        will  be  displayed  in  the  lower  right 

        corner  of  the  LCD.

O

C

CON

TEMP

MEMO

uS

ATC

LCD  DISPLAY

PH  probe

Содержание 850048

Страница 1: ...DO Meter Kit 850048 Instruction Manual...

Страница 2: ...9 When using the meter press the buttons firmly for one full second to ensure proper operation DO Meter Kit 850048 Copyright 2013 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents...

Страница 3: ...Clear P10 12 DO Review and Set Parameters P60 13 Select Temperature Units P70 14 Restore Default Settings P90 15 pH Electrode Slope P20 16 Conductivity Calibration P30 17 Conductivity Cell Constant P...

Страница 4: ...F multiple calibration points for optional pH and conductivity probes 99 memory points a large backlit LCD screen a low battery icon and one hour automatic shut off A Set Up Mode allows you to custom...

Страница 5: ...into the connector port on top of the meter 2 Screw down the attached probe guard until it is moderately tight To detach the probe from the meter 1 Unscrew the probe guard the piece that is directly...

Страница 6: ...CON s or Salt ppt units RECALL Press for 2 seconds to read the memory value mg L Press to switch between DO in and mg l Press to increase the displayed value in Function or Calibration Modes MEM Pres...

Страница 7: ...value will be isplayed in the upper right corner of he LCD The MEMO number which is the total umber of records saved in the meter will be displayed in the lower right orner of the LCD DD O probe DO PR...

Страница 8: ...s ENTER to finish the calibration 6 If ERR appears during calibration there is an error due to low electrolytes or a defective probe Note To exit Calibration Mode without saving the new value press CA...

Страница 9: ...esired by pressing mg L 6 Place the probe into the sample solution making sure that the electrode is completely immersed in the sample 7 Stir the probe gently to create a homogenous sample 8 Wait unti...

Страница 10: ...ALL for 2 seconds to enter Recall Mode 2 Press or to scroll through the data in memory 3 Press CAL ESC for 2 seconds to return to Normal Mode O C DO TEMP MEMO mg l ATC Fig 18 Press any key to actirate...

Страница 11: ...right side shows the function number 3 Press CAL ESC to exit Function Mode The table below shows the programmable functions of the DO probe as well as the optional pH and Conductivity probes Paramete...

Страница 12: ...stored with the DO probe 1 Enter Function Mode Clr P10 will be displayed on the LCD screen 2 Press ENTER The default no icon will flash and P11 will be displayed 3 Press or to change the status from...

Страница 13: ...pt O C DO DO DO M 14 Press Enter to confirm then enter P63M to measure height from sea level with manually enter compensation Fig13 The value will be flashed on the screen You can manually enter or ke...

Страница 14: ...on the left side of the screen Fig 14 2 At P70 screen press ENTER key to enter P71 The default C will be flashed on the main screen Fig15 If the unit you need is degree o C press ENTER key to confirm...

Страница 15: ...key to confirm The screen will return to P90 screen Fig 16 Fig 16 Fig 17 16 Parameters Default Setting Value P10 Memory Clear CLr adjustable No P20 PH Electrode Slope Review 100 P30 Cond Calibration...

Страница 16: ...the ELE P20 Fig 4 display 2 Press ENTER to view P22 the slope value will be displayed on LCD same as P23 Fig 5 As maximum 3 PH calibration points is allowed in this meter so the user can review two sl...

Страница 17: ...information in Range 4 Press ENTER key to return to CAL 30 screen Fig 6 Fig 6 Fig 7 CON CON uS 12 Press or key in the setup mode to check the CAL P30 calibration message Fig 6 This feature is only to...

Страница 18: ...the conductivity probe Press or to adjust the salinity value The value can be adjusted from 0 0 to 42 00ppt CON CON Press or key in the setup mode to check CEL P40 display Fig 8 Press ENTER key to en...

Страница 19: ...ht moderately Parameters Default Setting Value P 10 Memory Clear CLr adjustable No P 20 pH Electrode Slope Review 100 P 30 Cond Calibration Review 146 6 S 1413 S 12 88 mS 51 5 mS P 40 Cond Cell Consta...

Страница 20: ...e set 4 Put the probe protective cap back 5 Screw the probe protective cap tightly To refill DO electrolyte solution Refilling the Membrane Set Electrolyte Solution Replace the membrane set when the m...

Страница 21: ...ity when you insert the batteries TROUBLESHOOTING Meter is on but there is no display Check and or replace the batteries Be sure there is good contact and that the polarity is correct Meter is not rea...

Страница 22: ...l data Check the mV or temperature reading E13 There is a mistake in the pH calibration pH probe only Recalibrate or replace the probe E16 The cell constant is out of range COND probe only Recalibrate...

Страница 23: ...m 1 F S 1 digit 0 1 S cm 1 S cm 0 01 mS cm 0 1 mS cm Salinity 0 10 ppt 0 42 ppt sea water 1 F S 1 digit 0 01 ppt 0 1 ppt DO 0 199 9 0 30 mg L 3 F S 1 digit 0 1 Temp 0 60 0 C 0 5 C 0 1 Operating Temp R...

Страница 24: ...24...

Страница 25: ...25 Appendix A 850048PH pH Probe...

Страница 26: ...pH probe 2 Rinse the electrode in de ionized or distilled water If the probe is dehydrated soak it for 30 minutes in a KCI solution 3 Insert the pH probe into the meter 4 Press the power button to tu...

Страница 27: ...meter using 4 01 10 01 buffers O C PH CAL O C PH 4 Press CAL ESC key to enter calibration mode 5 Once the probe auto recognize the pH buffer For example 7 01 the CAL and 7 01 will flash Fig 19 6 If th...

Страница 28: ...wipe or rub the glass bulb The pH glass bulb must always be kept moist Place it in the protective storage cap whenever it is not in use Press to confirm the calibration or parameter setting LCD DISPLA...

Страница 29: ...29 Appendix B 850048CS Conductivity Salinity Probe...

Страница 30: ...ions For example if the measuring range is 0 1999 mS use 1413 S solution to calibrate You will only need to do single point calibration Calibration Procedure 1 Remove the protective cap from the condu...

Страница 31: ...the probe on the bottom of the container to remove any air bubbles from the sensing area 8 Press and hold CAL ESC for 2 seconds to enter Calibration Mode The probe will detect the conductivity soluti...

Страница 32: ...rn to normal mode then the conductivity calibration is completed 21 Fig 22 NOTE The characteristics of the Probe may degrade with time and usage Whenever an E16 occurs after calibration please replace...

Страница 33: ...eter at least once a week if measurement is taking place at extreme temperatures CONDUCTIVITY MEASUREMENT 1 Remove the protective cap from the conductivity probe 2 Rinse the probe by inserting it into...

Страница 34: ...reading has stabilized CONDUCTIVITY PROBE MAINTENANCE Do NOT rub or wipe the surface of the electrode This might change the original constants and affect the testing range Protect the probe in the st...

Страница 35: ...35...

Страница 36: ...robes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof seal and void the warra...

Отзывы: