DO Review and Set Parameters Function (P60)
1. Enter Function Mode. Press ▲or ▼ until “COE P60”
is displayed.
2. Press
ENTER
to view the previous DO calibration
temperature information. “P61” will be displayed.
3. Press
ENTER
again. The salinity compensation
value will flash and “P62” will be displayed.
DO
ppt
O
C
DO
DO
DO
M
14
Press Enter to confirm. then enter
P63M to measure height from sea level
with manually enter compensation
(Fig13). The value will be flashed on
the screen. You can manually enter
△
or
▽
key to adjust from 0~3500M (100
M per step),press ENTER to confirm
and return to COE P60 screen.(Fig 10)
Fig.
10
Fig.11
Fig.12
Fig.13
.
DO
ppt
O
C
DO
DO
DO
M
14
Press Enter to confirm. then enter
P63M to measure height from sea level
with manually enter compensation
(Fig13). The value will be flashed on
the screen. You can manually enter
△
or
▽
key to adjust from 0~3500M (100
M per step),press ENTER to confirm
and return to COE P60 screen.(Fig 10)
Fig.
10
Fig.11
Fig.12
Fig.13
.
DO
ppt
O
C
DO
DO
DO
M
14
Press Enter to confirm. then enter
P63M to measure height from sea level
with manually enter compensation
(Fig13). The value will be flashed on
the screen. You can manually enter
△
or
▽
key to adjust from 0~3500M (100
M per step),press ENTER to confirm
and return to COE P60 screen.(Fig 10)
Fig.
10
Fig.11
Fig.12
Fig.13
.
Note…
To exit at any point without making any changes,
press
CAL/ESC
.
4. Press or hold down ▲or ▼ to adjust the salinity
compensation value. The salinity value can be
adjusted from 0.0 to 42.0 ppt.
5. Press
ENTER
to confirm.
6. The compensation value for height above sea level will
flash and “P63” will be displayed. Press or hold down
▲or ▼ to adjust the value. The height level can be
adjusted from 0 to 3500 m in 100 m increments.