background image

IMPORTANT: Please read this information and your TB100 user’s manual before 

using TB100 Test Strips. Call Spengler Customer Service immediately if these 

instructional materials or your meter results seem unclear.

Intended Use

The TB100 Blood Glucose Test Strips are used with the TB100 Blood Glucose 

Meter for quantitatively measuring glucose (sugar) in whole blood obtained 

from the fingertip, palm and forearm. The TB100 Blood Glucose Monitoring 

System is intended for use in the home and in professional settings to monitor 

blood glucose levels. The TB100 Test Strips are for testing outside the body (in 

vitro diagnostic use).

Test Principle

On each test strip there is a test area containing reaction chemicals. When blood 

is applied to this area, a chemical reaction takes place, then a transient electrical 

current is formed. The blood glucose concentration is calculated based on the 

electrical current detected by the meter, then the result is showed on the 

display. The test measures glucose from 20 mg/dL (1.1mmol/L) to 600 mg/dL 

(33.3 mmol/L). The TB100 Test Strip is calibrated to display the equivalent of 

plasma glucose values to allow easy comparison of results with laboratory 

methods.

Precautions

˙ 

For in vitro diagnostic use (outside the body) only.

˙

Do not reuse test strips.

˙

Do not use test strips beyond the expiration date printed on the package since 

    this may cause inaccurate results.

˙

Discard the vial and any unused test strips 3 months after you first open it. 

    Constant exposure test strips to air may destroy chemicals in the test strip. This 

    damage can cause incorrect readings.

˙

The sample must only be applied to the tip of the test strip. Do not apply blood 

    or control solution to the top, left or right sides of the test strip. This may cause 

    inaccurate test results.

˙

Use only the TB100 Blood Glucose Meter with the TB100 

    Test Strips.

˙

Keep the test strip vial away from children. The cap is a choking hazard. The 

    cap or vial contains drying agents to protect the test strips. Drying agents may 

    be harmful if inhaled or swallowed and may cause skin or eye irritation.

˙

Any change in medication based on the TB100 blood glucose 

    test results without the consent and advice of a physician or healthcare 

    professional is not recommended. Consult your physician or healthcare 

    professional before making changes in your treatment plan based on 

    your blood glucose test results.

˙

Severe dehydration and excessive water loss may cause false low results. If you 

    believe you are suffering from severe dehydration, consult your physician 

    immediately.

˙

All devices contaminated with blood should be disposed of properly.

˙

If you get TB100 Control Solution test results that fall outside the 

    range printed on the test strip vial, repeat the test with a new test strip. If the 

    test result is still outside the range printed on the test strip vial, the TB100 may 

    not be working properly. DO NOT use the system to test your blood 

    until you get a control solution test result within the range printed on the test 

    strip vial label.

Storage and Handling

˙

Store at room temperature between 4-30

 (39-86

). Do not refrigerate or 

    freeze. Avoid exposing test strips to extreme temperatures.

˙

Store away from direct sunlight and heat.

˙

Use each strip immediately after removing it from the vial.

˙

Store your test strips in their original vial only. The cap or vial contains drying 

    agents to protect the test strips. Do not transfer test strips to a new vial or any 

    other container.

˙

After removing a test strip from the vial, replace the vial cap immediately and 

    close it tightly.

˙

Do not bend, cut, or alter the Test Strip in any way.

˙

With clean, dry hands you may gently touch the test strip anywhere when 

    removing it from the vial or inserting it into the meter.

 

How to do the Test

See your User’s Manual for a step-by-step guide on how to do the test.

Warning

Alternate site testing (AST) should ONLY be used in the following intervals:

˙

In a pre-meal or fasting state. (more than 2 hours since the last meal)

˙

Two hours or more after taking insulin.

˙

Two hours or more after exercise.

DO NOT use AST if: 

˙

You think your blood glucose is low.

˙

You are unaware of hypoglycemia.

˙

Your AST results do not match the way you feel.

˙

You are testing for hyperglycemia.

˙

Your routine glucose results are often fluctuating.

˙

If you are pregnant.

All parts of the glucose monitoring system should be considered potentially 

infectious and are capable of transmitting blood-borne pathogens between 

each user. To assure that you are not placing yourself or other persons at risk, 

always remember:

˙

The meter and lancing device are for single person use.

    Do NOT share them with anyone including other family members! Do not use    

    on multiple persons!

˙

All parts of the kit are considered biohazardous and can potentially transmit 

    infectious diseases, even after you have performed cleaning and disinfection.

˙

Refer to your user’s manual section “Cleaning and Disinfection procedure” for 

    how to clean your meter and lancing device.

1.

2.

3.

What Does The Results Mean?

Normal Glucose Values

The normal fasting glucose range for a diabetic adult is 70 to 130 mg/dL (3.9 to 
7.2 mmol/L). One to two hours after meals, normal glucose values should be less 
than 180 mg/dL (10.0 mmol/L).1 For people with diabetes, consult your 
physician or healthcare professional for the target glucose values that are right 
for you. 

Low Glucose Values

The TB100 Meter displays results between 20 and 600 mg/dL (1.1 and 33.3 
mmol/L). If your test result is lower than 20, “Low” (LO) will appear on the meter 
display. This indicates severe low blood sugar (hypoglycemia). You should 
immediately treat low blood sugar as recommended by your healthcare 
professional.

High Glucose Values

If your test result is above 600 mg/dL (33.3 mmol/L), “High” (HI) will appear on 
the meter display screen. This indicates severe high blood sugar (hyperglyce-
mia). You should immediately treat high blood sugar as recommended by your 
healthcare professional.

Unexpected Results

Low or high blood sugar readings can indicate a potentially serious medical 
condition. If your blood sugar is unusually low or high, or if you do not feel the 
way your results indicate, repeat the test with a new test strip.  

If your reading is 

not consistent with your symptoms or if your blood glucose result is less than 60 
mg/dL (3.3 mmol/L) or higher than 240 mg/dL (13.3 mmol/L), you should contact 
your healthcare professional and follow his or her treatment advice.

Limitations
The TB100 Blood Glucose Test Strips give accurate results when the following 
limitations are observed:
  1.

  2.

  3.

 
  4.

  5.
  6. 
  7.

  8.

  9. 

10.
 
11.

12.

Use only the TB100 Test Strip with the TB100 Meter.
Use only capillary whole blood from finger, palm and forearm. Do not use 
plasma or serum.
Do not use neonate samples. TB100 Test Strips are not validated for and 
should not be used for testing neonatal blood specimens.
The system is tested to accurately read the measurement of glucose in 
whole blood within the range of 20 to 600mg/dL (1.1-33.3 mmol/L).
This system can be used up to an altitude of 7545 feet.
The test strips are for single use only. Do not reuse test strips.
TB100 Test Strip does not interfere with the hematocrit at a normal range 
(35-55%) of blood glucose. High (above 55%) and low (below 35%) 
hematocrit can cause false results. Patients with higher hematocrit values 
(above 55%) will usually test lower whiles patients with lower hematocrit 
values (below 35%) will test higher. Please ask your healthcare professional 
if you do not know your Hematocrit level.
Acetaminophen, uric acid, ascorbic acid (vitamin C), and other reducing 
substances when occurring in normal blood or normal therapeutic 
concentrations do not significantly affect results. However, abnormally high 
concentrations in blood may cause inaccurately high results. Lipemic 
samples; Cholesterol up to 375 mg/dL or triglycerides up to 2,000 mg/dL do 
not significantly affect test results. However, glucose values in specimens 
beyond these levels, should be interpreted with caution.
Blood samples that contain a high concentration of dissolved oxygen may 
lower the test result.
Antiglycolysis and anticoagulants in blood samples may affect the test 
results.
Severe dehydration, diabetic ketoacidosis, hyperosmolar non-kitotic state, 
hypotension, shock, or peripheral vascular disease.2,3,4 may cause 
inaccurate results.
Critically ill patients should not be tested with home use blood glucose 
monitors as it may lead to inaccurate results.

Select the Puncture Site to Obtain Blood
You may obtain a blood drop from the fingertip, palm, or forearm. If using the 
forearm, choose a fleshy area away from bone, visible veins and hair. If using 
the palm, choose a fleshy area on the palm below your thumb or pinky finger. 
Select a spot with no visible veins and away from deep lines, which may cause 
your blood sample to smear. Wash your hands and the puncture site with 
warm, soapy water. Dry them thoroughly. If you use an alcohol swab, make 
sure the skin is dry before lancing. To help you get a large enough drop of 
blood, you may gently massage or apply warmth to the site to increase blood 
flow. Press and hold the sampler against your fingertip, palm or forearm for a 
few seconds, then press the release button.
Blood Glucose Testing
Insert a Test Strip to the meter. The meter will turn on. Use the Lancing Device 
to obtain the right size blood drop. Refer to your Lancing Device User’s 
Instruction for how to lance your finger, forearm and palm. Then obtain a 
blood sample about the size of a pinhead. Gently touch only one edge of the 
Test Strip to the blood sample. Do not press the edge of the strip against the 
test site. When the strip is full of blood, the meter will “beep” or you will see 
the test strip chamber area completely filled.
Read Results
Read the test results in an average of 5 seconds. The result will be stored in 
the meter memory. Results are displayed as mg/dL (milligrams per deciliter of 
blood) or mmol/L (millimoles per liter of blood) depending on the units you 
chose when you set up your meter. To change the units displayed, see your 
User’s Manual section “Setting Up the Meter.” 

Wash hands thoroughly with soap and water after handling the meter, 

lancing device, or test strips.

1

For Testing Glucose in Capillary Whole Blood

TB100 Test Strip Package Insert

Содержание TB100

Страница 1: ...cémie Les valeurs normales de glycémie pour un adulte diabétique à jeun sont comprises entre 70 et 130 mg dL entre 3 9 et 7 2 mmol L Une à deux heures après un repas les valeurs normales de glycémie doivent être inférieures à 180 mg dL 10 0 mmol L 1 Les personnes diabétiques doivent consulter leur médecin ou tout autre professionnel de santé afin de définir la plage de valeurs qui s applique à ell...

Страница 2: ......

Страница 3: ...s The normal fasting glucose range for a diabetic adult is 70 to 130 mg dL 3 9 to 7 2 mmol L One to two hours after meals normal glucose values should be less than 180 mg dL 10 0 mmol L 1 For people with diabetes consult your physician or healthcare professional for the target glucose values that are right for you Low Glucose Values The TB100 Meter displays results between 20 and 600 mg dL 1 1 and...

Страница 4: ......

Страница 5: ... de glucosa normal en ayuno para un diabético adulto va de 70 a 130 mg dL 3 9 a 7 2 mmol L Una o dos horas después de las comidas los valores de glucosa normales deberían ser inferiores a 180 mg dL 10 0 mmol L 1 Para personas con diabetes consulte a su médico o profesional sanitario para saber los valores de glucosa objetivo adecuados para usted Valores de glucosa bajos El medidor TB100 muestra un...

Страница 6: ......

Страница 7: ...ﺗر ﻣﻠﺟم 180 ﻣن أﻗل اﻟطﺑﯾﻌﯾﺔ اﻟﺟﻠوﻛوز ﻗﯾم ﺗﻛون أن ﯾﺟب ﻟﺗر ﻣﻠﯾﻣول 7 2 إﻟﻰ اﻟﺳﻛري ﺑداء اﻟﻣﺻﺎﺑﯾن ﻟﻸﺷﺧﺎص ﺑﺎﻟﻧﺳﺑﺔ 1 اﻟوﺟﺑﺎت ﺗﻧﺎول ﻣن ﺳﺎﻋﺗﯾن إﻟﻰ ﺳﺎﻋﺔ ﺑﻌد ﻟﺗر ﻣﻠﯾﻣول ﻟﻛم اﻟﻣﻧﺎﺳﺑﺔ اﻟﻣﺳﺗﮭدﻓﺔ اﻟﺟﻠوﻛوز ﻗﯾم ﻟﻣﻌرﻓﺔ اﻟﺻﺣﯾﺔ اﻟرﻋﺎﯾﺔ أﺧﺻﺎﺋﻲ أو طﺑﯾﺑﻛم اﺳﺗﺷﯾروا اﻟﻣﻧﺧﻔﺿﺔ اﻟﺟﻠوﻛوز ﻗﯾم ﻛﺎﻧت إذا ﻟﺗر ﻣﻠﯾﻣول 33 3 و 1 1 دﯾﺳﯾﻠﺗر ﻣﻠﺟم 600 و 20 ﺑﯾن اﻟﻧﺗﺎﺋﺞ TB100 اﻟﻘﯾﺎس ﺟﮭﺎز ﯾﻌرض إﻟﻰ ھذا ﯾﺷﯾر اﻟﻘﯾﺎس ﺟﮭﺎز ﺷﺎﺷﺔ ﻋﻠﻰ ﻣﻧﺧﻔض LO Low ﻛ...

Страница 8: ......

Отзывы: