Oxystart® - EN
Guidance and manufacture’s declaration-electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic immunity
The Oxystart
®
Pulse Oximeter is intended for use in the electromagnetic environment specified below. The
user of the Oxystart
®
Pulse Oximeter should ensure that it is used in such an environment.
Immunity test
IEC60601 test
level
Compliance level Electromagnetic
environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6KV contact
±8KV air
± 6 kV contact
± 8 kV air
Floors should be made of wood, concrete or
ceramic tile. If the floor is covered with syn-
thetic material, the relative humidity should
be at least 30%.
Power frequency
(50Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment