3.4 

Technical data  

Lifting bridles system 

STX 519 

STX 540 

STX 542 

Webbing length (mm) 

maximum 1680 

minimum 1250 

1600 and 1550 

1660 

Webbing width (mm) 

50 / 20 

30 

30 

Weight (kg) 

1,8 

0,8 

0,6 

Material 

nylon, aluminium, steel 

nylon, aluminium, steel 

nylon, aluminium, steel 

Type of fixation 

carabineer 

noose 

carabineer 

Number of fixing straps 

Compatibility 

all basket stretchers 

made by Spencer 

all basket stretchers made by 

Spencer 

all basket stretchers made by 

Spencer 

Maximum static safety load 

450 kg (1000 lbs) 

450 kg (1000 lbs) 

450 kg (1000 lbs) 

 
3.5 Reference standards 

Reference 

Title of document 

MDD 93/42/CEE 

European Directive about Medical Devices 

MDD 2007/47/CEE 

Modifications to 90/385/CEE Directive about active implants, Directive 
93/42/CEE  about  medical  devices  and  Directive  98/8/CE  about  the 
introduction of biocides onto the market 

Legislative Decree 24/02/1997, n. 46 

Application of the 93/42/CEE Directive about Medical Devices 

Legislative Decree 25/01/2010, n. 35 

Modifications and additions to the 20/02/97 Decree n. 46 

UNI EN ISO 14971 

Application of risks managing to medical devices 

UNI CEI EN ISO 15223-1 

Graphic symbols used for medical devices labelling 

UNI CEI EN 1041 

Information supplied by the medical devices manufacturer 

CEI EN 62366 

Medical  Devices  -  Application  of  the  utilisation  characteristics  of 
engineering to medical devices 

MEDDEV 2.4/1a-b 

Guideline for the classification of medical devices 

NB-MED 2.5.1 /Rec 5  

Technical Documentation 

MEDDEV 2.7.1  

Clinical Data 

MEDDEV 2.12/1  

Medical Devices vigilance system 

UNI EN 14155 

Clinical  evaluation  of  the  medical  devices  for  human  beings  -  Part  2: 
Clinical evaluation plans 

 
3.6 

Environmental conditions 

Functioning temperature: from -20 to +60 °C 
Storage temperature: from -20 to +60 °C 
Relative humidity: from 5 to 85% 
 

4. 

OPERATING INSTRUCTIONS 

4.1 

Transport and storage 

Before transporting the appliance, make sure that it is correctly packaged ensuring also that there are no risks of  damages 
during the transport itself. 
Keep the original packaging for use in case of any further transport and for storage. Damage to the appliance caused during 
transport and handling is not covered by the guarantee. Repairs or replacement of the damaged parts are the responsibility 
of the client. The device must be stored in a dry, cool area away from direct sunlight. It must not be placed in contact with 
any substances or chemical agents which could cause damage and reduce safety characteristics.

 

 

4.2 

Preparation 

On receipt of the product: 

 

Remove the packaging and display the material so that all components are visible. 

 

Check that all the components/pieces on the accompanying list are present. 

The  appliance  must  be  checked  before  every  use  so  as  to  reveal  any  working  abnormalities  and/or  damage  caused  by 
transport and/or storage. In particular, check: 

 

General functionality of the device 

 

Cleanliness of the device (remember that the failure of cleaning may cause the risk of cross infections) 

 

Absence of cuts, holes, tears on the structure, including the straps. 

 

Do the straps fold in the correct way? 

 

Correct grasp of the straps 

 

Status of wear of the components (straps, buckles) 

 

Proper working of buckles  

Содержание STX 519

Страница 1: ...rol of the products certified by the notifying body TÜV SÜD Product Service GmbH INDEX General information page 2 Warning page 3 Description of product page 4 Operating instructions page 6 Maintenance and cleaning page 7 Accessories and spare parts page 8 First edition 07 02 02 Rev 9 18 12 17 Spencer Italia s r l Tel 39 0521 541111 Fax 39 0521 541222 e mail info spencer it STX 519 STX 540 STX 542 ...

Страница 2: ...and or on the box This label includes information about the manufacturer the product CE mark lot number LOT It must never be removed or covered 2 WARNINGS 2 1 General warnings The product must be used by trained personnel only having attended specific training for this device and not for similar products Training routines must be registered on a special register in which the names of those trained...

Страница 3: ... to report any accident that involves any medical device to the Ministry of Health and to the Manufacture as specified and within time given by the European regulations In addition both public and private operators are obliged to inform the manufacturer of any measures that should be adopted to make the steps necessary to guarantee the safety and the health of the patients and the users o any medi...

Страница 4: ...ling the device on rough surfaces The device lifting device used must be approved by the actual norm The user is responsible to evaluate the application and the risks connected with the use of the device Always check the integrity of all the parts of the straps and buckles before every use Immediately replace straps and buckles if they are worn out or damaged Fix the straps and buckles into the de...

Страница 5: ...1 Clinical Data MEDDEV 2 12 1 Medical Devices vigilance system UNI EN 14155 Clinical evaluation of the medical devices for human beings Part 2 Clinical evaluation plans 3 6 Environmental conditions Functioning temperature from 20 to 60 C Storage temperature from 20 to 60 C Relative humidity from 5 to 85 4 OPERATING INSTRUCTIONS 4 1 Transport and storage Before transporting the appliance make sure ...

Страница 6: ...nism is broken Immediately remove the device from service and contact the service centre 5 MAINTENANCE AND CLEANING 5 1 Cleaning Failure to carry out the correct cleaning routine could increase the risk of cross infection The operator must always wear adequate personal protection such as gloves and mask etc during all checking and cleaning procedures The metal parts exposed to external agents unde...

Страница 7: ...lity of the adjustment system Check the presence and legibility of the label containing warning data life span and loading capacity Depending on the intensity of use the device must undergo revision procedures to certify its suitability for use during the permitted life span of the product according to par 5 2 4 Such revisions shall be formalized and documented in a report with certain date which ...

Страница 8: ...8 6 ACCESSORIES AND SPARE PARTS 6 1 Accessories There aren t any accessories for these items 6 1 Spare parts There aren t any spare parts for these items ...

Страница 9: ... having attended specific training for this device and not for similar products Keep this document at least 10 years from the end of life of the device Operator s name Training date Training method user s manual during service former class etc Trainer Basic training Advanced training ...

Страница 10: ...ified without any warning and is not to be intended as a commitment on behalf of Spencer Italia S r l Spencer products are exported to many countries and the same identical regulations are not always valid For this reason there could be differences between the description here described and the product actually delivered Spencer continually strives to reach the perfection of all items sold We ther...

Страница 11: ... the device under test The load shall be applied simulating the weight distribution of the human anatomy according to the following table 450 KG 9 HEAD AND NECK 39 CHEST AND ARMS 26 PELVIS 16 THIGHS 10 LEGS The simulation shall be carried out lifting the load slowly avoiding sudden application of the load Maintain the maximum load condition for 30 minutes It is therefore necessary to fill in the f...

Страница 12: ...r testing pictures Basket strether pictures Bridles picture Picture of wights used for tests PICTURES AFTER MAXIMUM LOAD TEST CARABINERS FILL WITH PICTURES PROVING THE INTEGRITY OF CARABINERS BELTS FILL WITH PICTURES PROVING THE INTEGRITY OF BELTS SEAMS FILL WITH PICTURES PROVING THE INTEGRITY OF SEAMS ...

Страница 13: ...true to their original form and are free of cracks deep scratches and gouges All belts and seams are in optimal condition and with no cuts or frayed edges Following the tests and visual inspection documented above the test s responsible declares that having not exceeded its life time the device is This certificate has a validity of 1 year Place Date Test s responsible Signature NOT SUITABLE FOR US...

Отзывы: