Main Input Power Connections (I/O System Interface Box):
The main input power connections will depend on the power rating of the XRV Generator. A 4kW XRV
Generator will require a single phase 180Vac to 264Vac source while the 6kW XRV Generator requires a
3 phase /-10% Input. The input voltage requirement for the 6kW units must be specified by the
user when purchased. If the user installs the input power cables, the appropriate wire size must be used.
Recommended wire for the 4kW generators and below is copper stranded #8 AWG. The wire size for the
6kW 3 phase systems is #10AWG due to the higher current draw on the input line. With the exception of
the Cooler/Chiller and Power Door Interlock connections, all of the cabling and connectors are provided
with the I/O System Interface Box to easily connect all of the subsystem components.
Note: Although a 3 phase 400Vac system can be selected, the I/O Subsystem Box for that voltage is
not available at this time.
Содержание XRV Sub-system
Страница 1: ...SPELLMAN HIGH VOLTAGE XRV SUBSYSTEM INSTALLATION MANUAL ...
Страница 13: ...Page 13 of 98 118157 001 Rev C XRV 160kV Overview 400Vac 3Ø Input ...
Страница 17: ...Page 17 of 98 118157 001 Rev C XRV225kV Overview XRV225 400Vac 3Ø 6kW Unit ...
Страница 21: ...Page 21 of 98 118157 001 Rev C XRV450 Overview XRV450 Bipolar Cathode Generator ...
Страница 22: ...Page 22 of 98 118157 001 Rev C XRV450 Overview XRV450 Bipolar Anode Generator ...
Страница 25: ...Page 25 of 98 118157 001 Rev C I O System Interface Box Overview I O System Interface Box Front Panel Detail ...
Страница 28: ...Page 28 of 98 118157 001 Rev C XRVC Controller Overview XRVC Controller Rear Panel Detail ...
Страница 34: ...Page 34 of 98 118157 001 Rev C XRV Unipolar System Diagram Typical Unipolar System Minimal Requirements ...
Страница 36: ...Page 36 of 98 118157 001 Rev C XRV Bipolar System Diagram Typical Bipolar System Minimal Requirements ...
Страница 51: ...Page 51 of 98 118157 001 Rev C System Grounding Detail and HV Cable Connections ...
Страница 58: ...Page 58 of 98 118157 001 Rev C Interlock Communication Connections Unipolar Basic RS 232 Cable ...
Страница 59: ...Page 59 of 98 118157 001 Rev C Interlock Communication Connections Unipolar Basic Ethernet Cable USB Cable ...
Страница 66: ...Page 66 of 98 118157 001 Rev C ...
Страница 67: ...Page 67 of 98 118157 001 Rev C Power Source Connections Unipolar Basic ...
Страница 68: ...Page 68 of 98 118157 001 Rev C Power Source Connections Unipolar Basic ...
Страница 72: ...Page 72 of 98 118157 001 Rev C Power Connections I O System Interface Box ...
Страница 76: ...Page 76 of 98 118157 001 Rev C Main Input Power Connections I O System Interface Box ...
Страница 77: ...Page 77 of 98 118157 001 Rev C I O Subsystem Interface Box System Diagram ...
Страница 82: ...Page 82 of 98 118157 001 Rev C Flash Drive USB installation Step 1 Flash Drive USB installation Step 2 ...
Страница 83: ...Page 83 of 98 118157 001 Rev C Flash Drive USB installation Step 3 Flash Drive USB installation Step 4 ...
Страница 89: ...Page 89 of 98 118157 001 Rev C X Ray Tube Filament Calibration Setup ...