1.4.3.8.
Request
ADC
Data
(20)
This
command
requests
that
the
unit
returns
various
converted
analog
values.
1.
Positive
KV
(
3983
equal
to
unit
full
scale
in
kilovolts)
2.
mA
(3983
equal
to
unit
full
scale
in
milliamps)
3.
Not
used
4.
+24V
(3246
equal
to
24V)
5.
+15V
(2028
equal
to
15V)
6.
+10V
(1353
equal
to
10V)
7.
+5V
(3412
equal
to
5V)
8.
+3.3V
(2252
equal
to
3.3V)
9.
Negative
KV
(3983
equal
to
unit
full
scale
in
kilovolts)
Example:
Sent
to
Unit
<STX>20,<ETX>
Received
from
unit
<STX>20,0,1,4,3223,1867,1407,3417,2255,0,<ETX>
1.4.3.9.
Enable
High
Voltage
(99)
This
commands
turns
high
voltage
on
if
there
is
no
faults
and
interlock
is
closed.
No
arguments
required.
Example:
Sent
to
Unit
<STX>99,<ETX>
Received
from
unit
<STX>99,$,<ETX>
1.4.3.10.
Disable
High
Voltage
(98)
This
command
turns
high
voltage
off.
No
arguments
required.
Example:
Sent
to
Unit
<STX>98,<ETX>
Received
from
unit
<STX>98,$,<ETX>