8-2
Chapter 8: Verifying Calibration
User’s Manual Pub. 0300128-04 Rev. A
3.
Use the precision input source to provide a 15 mA input across a
channel’s input terminals.
4.
Verify that the input data is between 15110 and 15260 counts.
5.
Repeat steps 3 and 4 for each remaining channel.
Section 8.3
Verifying Voltage
Inputs (INI4vi only)
To verify:
1.
Configure your module for ±10 V input on all 4 channels using the
following configuration word values:
Address
Binary Value
Decimal Equivalent
O:e.0
0000 0000 0010 0001
33
O:e.1
0000 0000 0010 0001
33
O:e.2
0000 0000 0010 0001
33
O:e.3
0000 0000 0010 0001
33
e = slot number
2.
Let the operating temperature stabilize for 2 minutes.
3.
Use the precision input source to provide a 10 VDC input across a
channel’s input terminals.
4.
Verify that the input data is between 31805 and 32131 counts.
5.
Repeat steps 3 and 4 for each remaining channel.
Section 8.4
Service
If the input data is out of range, call the Spectrum Controls Technical Services to
make arrangements for returning your module for factory calibration.
Содержание 1746sc-INI4i
Страница 1: ...User s Manual Pub 0300128 04 Rev A...
Страница 8: ...viii SLC 500 Isolated Analog Input Modules User s Manual Pub 0300128 04 Rev A...
Страница 14: ...1 6 Chapter 1 Module Overview User s Manual Pub 0300128 04 Rev A...
Страница 30: ...3 8 Chapter 3 Configuring the 1746sc INI4I User s Manual Pub 0300128 04 Rev A...
Страница 44: ...5 4 Chapter 5 Programming Examples User s Manual Pub 0300128 04 Rev A...
Страница 54: ...User s Manual Pub 0300128 04 Rev A...
Страница 55: ...User s Manual Pub 0300266 04 Rev A...