TROUBLESHOOTING
If the transilluminator does not yield satisfactory results, follow the
procedures below before attempting to service the unit:
a. Review and comply with the instructions in the Operation section.
b. Check that the transilluminator is connected to a power source
and that the on/off switch or timer knob is positioned correctly.
c. Check that the fuse is operating properly. Use needlenose pliers
to remove the fuse drawer from the power entry unit. There is an
extra fuse in the 120V unit. Remove and discard the spent fuse
and replace it with the spare. The 230V unit has two active fuses
and a filter. Both fuses must be replaced. Use metric fuses for both
the 120V (2.5A F) and 230V (1.5A F) units.
If these steps do not yield satisfactory results, refer to the following
chart:
Symptom
Probable Cause
Cure
9
1. Low output from part
of filter area
UV tube(s) functioning
at substandard
intensity level
Replace tube(s)
2. Low uniform output
(short wave or
medium wave
transilluminators only)
a. Filter solarized
a. Replace filter
assembly
b. UV tubes functioning
at substandard
intensity level
b. Replace tubes
c. Diffusing screen
solarized
(312nm and 365nm
models only)
c. Replace diffus-
ing screen
ENVIRONMENTAL CONDITIONS
The Spectroline fixed-intensity ultraviolet transilluminators are
designed to be safe under the following conditions.
• Indoor use;
• Altitude up to 2,000 m (6,562 ft.);
• Temperature 5°C to 40°C (41°F to 104°F);
• Maximum relative humidity 80% for temperatures up to 31°C
(88°F) decreasing linearly to 50% relative humidity at 40°C (104°F);
• Mains supply voltage fluctuations not to exceed ±10% of the
nominal voltage;
• Installation Category II;
• Pollution Degree 2.