8

6.3 LANGUAGE OPTIONS

All interactive features and displays of the AccuPRO 
and AccuPRO Plus are available in English, Spanish, 
German, French, and Chinese. Enter the SETTINGS 
menu and select the LANGUAGE option to enter the 
selection menu displayed below. Upon selection of the 
desired language, the language menu will refresh and 
appear in the newly selected language format.

NOTE:

 Should the language accidentally be changed 

to one which you cannot understand, the first (highest) 
option on the menu will revert to English as seen 
below.

Below is the same screen viewed in Chinese. The 
numbers remain the same. In this event, press  
UP until reaching option 1 then MENU to return to 
English. The same is applicable to the other 
languages, with the first option reverting to English.

7. INFORMATION

The INFORMATION screen, accessed via the third 
selection on the main menu, displays briefly when the 
device is powered on, and can be viewed for as long 
as desired through the menu option.

Important data including the serial number of your unit, 
date of last calibration, and software version is avaiable 
on the INFORMATION screen. This information will 
differ between the example below and your own unit.

To return to OPERATION Mode, select the BACK 
option.

8. CALIBRATION

The fourth and final option on the Main Menu is 
Calibration. Calibration requires a precise configuration, 
and cannot be performed with the device alone. Contact 
customer service at 1-516-333-4859 if you believe your 
AccuPRO may require calibration.

There is no need to enter this menu under normal 
circumstances. However, should you arrive on the 
calibration screen shown below, simply press MENU 
repeatedly to exit. 

The languages and their corresponding numbers 
match across all language settings. If you accidentally 
happen to select a language difficult to understand 
(for instance, Chinese), select Option Number 1 to 
revert back to the default English language setting. 
The same process is applicable to the other languages 
in the selection.

The INFORMATION Screen can be accessed in the Main 
Menu. By default, the INFORMATION Screen briefly 
displays during startup. To closely examine the data on 
the INFORMATION Screen, select INFORMATION from 
the Main Menu.
The serial number of your device, the date of your 
device’s most recent calibration, and the software 
version installed on your device is available on the 
INFORMATION screen.

Содержание Spectroline AccuPRO XP-2000A

Страница 1: ...1 OPERATOR S MANUAL 7 20 AM18029 2 PRINTED IN U S A AccuPRO Photometer Radiometer www Spectro UV com 4 Dubon Ct Farmingdale NY 11735 866 230 7305 XP 2000A ...

Страница 2: ...SENSOR WAVELENGTHS RANGES 6 6 CUSTOMIZING SETTINGS 6 1 CHANGING UNITS 7 6 2 ADJUSTING BRIGHTNESS 7 6 3 LANGUAGE OPTIONS 8 7 INFORMATON 8 8 CALIBRATION 8 9 9 THEORY OF OPERATION 9 10 WARRANTY MAINTENANCE AND BATTERY 10 1 WARRANTY 10 10 2 PREVENTIVE MAINTENANCE 10 10 3 BATTERY CHARGING 10 2020 by Spectro UV All rights reserved No part of this publication may be reproduced or transmitted in any form ...

Страница 3: ... included 1 3 PRECAUTIONS The AccuPRO is carefully designed to prevent accidental shock to the operator However no engineering design can render safe an instrument that is used carelessly Therefore the directions presented in this manual must be read carefully before any measurements are made Failure to follow directions could result in injury The UV sensor is designed for use in regions of the sp...

Страница 4: ...use Altitude up to 2 000 m 6 562 ft Temperature 5 C to 40 C 41 F to 104 F Maximum relative humidity 80 for temperatures up to 31 C 88 F decreasing linearly to 50 rela tive humidity at 40 C 104 F Mains supply voltage fluctuations not to exceed 10 of the nominal voltage Installation Category II Pollution Degree 2 DIMENSIONS Part Length Width Depth Weight Readout Unit 6 in 15 2 cm 3 0 in 7 6 cm 1 0 i...

Страница 5: ...from the MAIN MENU screen to thoroughly examine information pertaining to your device see Section 6 3 After loading the unit will boot directly to Operation mode displaying visible and UV readings by default To change meter display settings press MENU Then select the SETTINGS option When highlighted press MENU to enter SETTINGS menu see SETTINGS section for details 4 To take a measurement enter Op...

Страница 6: ...K value stores the maximum luminosity recorded since the last reset Simply press the DOWN arrow in operational mode to reset the PEAK value to zero across visible and UV light readouts 5 4 SENSOR WAVELENGTHS RANGES The single sensor attached to the AccuPRO will display UV A 365nm readings as well as visible light 555nm In addition the AccuPRO Plus will also display blue light at the wavelength of ...

Страница 7: ...een Select the desired portion of the spectrum using MENU then use the arrow keys to select the desired units The following example displays the options for visible light The example below shows the available units for ultraviolet light Upon pressing MENU a brief confirmation screen will display e g Units set to Lux before returning to the main menu Simply press MENU again to return to Operation M...

Страница 8: ...n the INFORMATION screen This information will differ between the example below and your own unit To return to OPERATION Mode select the BACK option 8 CALIBRATION The fourth and final option on the Main Menu is Calibration Calibration requires a precise configuration and cannot be performed with the device alone Contact customer service at 1 516 333 4859 if you believe your AccuPRO may require cal...

Страница 9: ...applied voltage condition and low levels of incident light the p n junction will generate a current proportional to the light power incident on the active surface This photon induced current or photocurrent will divide between the diode parallel dynamic resistance and the parallel load resistance The dynamic resistance is normally a high value and is an inverse exponential function of forward volt...

Страница 10: ...ock If storing for an extended period place the AccuPRO in its carrying case 10 3 BATTERY SAFETY AND CHARGING 1 The AccuPRO radiometer must be charged overnight or for at least 8 hours when the battery is replaced or after a battery reset removal and re installation of the same battery 2 CAUTION Do not use alkaline batteries in the AccuPRO as they may explode causing damage to the circuitry and po...

Страница 11: ...11 ...

Страница 12: ...12 www Spectro UV com 4 Dubon Ct Farmingdale NY 11735 866 230 7305 ...

Отзывы: