5

3. BUTTONS AND CONTROLS

 

• 

The keypad on the readout meter is equipped with three pressure-sensitive 
buttons that provide easy access to all functions and features.

• 

Press and hold down the MENU (ON/OFF) button to turn the meter ON.

• 

The UP and DOWN arrows to the left and right of the MENU button provide 
access to all displays and features. Selected menu items appear highlighted in 
blue. Selected Menu items can be confirmed by pressing the MENU button once.

• 

Pressing the DOWN arrow (while in OPERATION MODE) will reset the PEAK 
value (highest recorded luminescence since last reset) for visible and UV light 
sources. 

• 

Press and hold the MENU button in OPERATION mode to turn the unit OFF.

• 

Highlighting and selecting the BACK button will return the user to the previous 
screen without making any changes.

4. OPERATION––QUICK GUIDE 

1.  To turn the meter ON, press and hold the MENU button for two seconds. It is recommended 

to fully charge the unit before use (see section 10.3). 

2.  The Spectroline

®

 loading screen will appear as the unit powers ON. Information pertaining to 

your device will briefly appear as well. Select INFORMATION from the MAIN MENU screen to 
thoroughly examine information pertaining to your device (see Section 6). 

3.  After loading, the unit will boot directly to Operation mode, displaying visible and UV readings 

by default. To change meter display settings, press MENU. Then select the SETTINGS option. 
When highlighted, press MENU to enter SETTINGS menu (see SETTINGS section for 
details).

4.  To take a measurement, enter Operation Mode. Shine a light source directly over the center 

of the sensor. The measurement will display on the interface screen.

5.  The meter instantly displays light level data for each wavelength (intensity/unit area).
6.  To turn off the meter, press and hold the MENU button for 2 seconds while in Operation mode. 

The display will then power off. Wait at least 30 seconds after the meter powers down before 
performing a RESTART (by pressing and holding MENU for 2 seconds). The AccuPRO will 
only power off from Operation mode, and not from the menus.

7.  To both conserve battery life and preserve the sensitivity of the sensor head, utilize the "Set 

Auto-Off" function under the "SETTINGS" menu. (See section 6).

DOWN

 Key:

Reset PEAK Value

UP

 Key:

Toggle VIS/UV

MENU

 Key: 

Adjust settings, confirm choices

Operation Mode for Visible and UV Light Sources

UV-A Reading

Visible Reading

Units

UV PEAK Value

Battery Indicator

VIS PEAK Value

Содержание Spectroline AccuPRO XP-2000A

Страница 1: ...1 OPERATOR S MANUAL 7 20 AM18029 2 PRINTED IN U S A AccuPRO Photometer Radiometer www Spectro UV com 4 Dubon Ct Farmingdale NY 11735 866 230 7305 XP 2000A ...

Страница 2: ...SENSOR WAVELENGTHS RANGES 6 6 CUSTOMIZING SETTINGS 6 1 CHANGING UNITS 7 6 2 ADJUSTING BRIGHTNESS 7 6 3 LANGUAGE OPTIONS 8 7 INFORMATON 8 8 CALIBRATION 8 9 9 THEORY OF OPERATION 9 10 WARRANTY MAINTENANCE AND BATTERY 10 1 WARRANTY 10 10 2 PREVENTIVE MAINTENANCE 10 10 3 BATTERY CHARGING 10 2020 by Spectro UV All rights reserved No part of this publication may be reproduced or transmitted in any form ...

Страница 3: ... included 1 3 PRECAUTIONS The AccuPRO is carefully designed to prevent accidental shock to the operator However no engineering design can render safe an instrument that is used carelessly Therefore the directions presented in this manual must be read carefully before any measurements are made Failure to follow directions could result in injury The UV sensor is designed for use in regions of the sp...

Страница 4: ...use Altitude up to 2 000 m 6 562 ft Temperature 5 C to 40 C 41 F to 104 F Maximum relative humidity 80 for temperatures up to 31 C 88 F decreasing linearly to 50 rela tive humidity at 40 C 104 F Mains supply voltage fluctuations not to exceed 10 of the nominal voltage Installation Category II Pollution Degree 2 DIMENSIONS Part Length Width Depth Weight Readout Unit 6 in 15 2 cm 3 0 in 7 6 cm 1 0 i...

Страница 5: ...from the MAIN MENU screen to thoroughly examine information pertaining to your device see Section 6 3 After loading the unit will boot directly to Operation mode displaying visible and UV readings by default To change meter display settings press MENU Then select the SETTINGS option When highlighted press MENU to enter SETTINGS menu see SETTINGS section for details 4 To take a measurement enter Op...

Страница 6: ...K value stores the maximum luminosity recorded since the last reset Simply press the DOWN arrow in operational mode to reset the PEAK value to zero across visible and UV light readouts 5 4 SENSOR WAVELENGTHS RANGES The single sensor attached to the AccuPRO will display UV A 365nm readings as well as visible light 555nm In addition the AccuPRO Plus will also display blue light at the wavelength of ...

Страница 7: ...een Select the desired portion of the spectrum using MENU then use the arrow keys to select the desired units The following example displays the options for visible light The example below shows the available units for ultraviolet light Upon pressing MENU a brief confirmation screen will display e g Units set to Lux before returning to the main menu Simply press MENU again to return to Operation M...

Страница 8: ...n the INFORMATION screen This information will differ between the example below and your own unit To return to OPERATION Mode select the BACK option 8 CALIBRATION The fourth and final option on the Main Menu is Calibration Calibration requires a precise configuration and cannot be performed with the device alone Contact customer service at 1 516 333 4859 if you believe your AccuPRO may require cal...

Страница 9: ...applied voltage condition and low levels of incident light the p n junction will generate a current proportional to the light power incident on the active surface This photon induced current or photocurrent will divide between the diode parallel dynamic resistance and the parallel load resistance The dynamic resistance is normally a high value and is an inverse exponential function of forward volt...

Страница 10: ...ock If storing for an extended period place the AccuPRO in its carrying case 10 3 BATTERY SAFETY AND CHARGING 1 The AccuPRO radiometer must be charged overnight or for at least 8 hours when the battery is replaced or after a battery reset removal and re installation of the same battery 2 CAUTION Do not use alkaline batteries in the AccuPRO as they may explode causing damage to the circuitry and po...

Страница 11: ...11 ...

Страница 12: ...12 www Spectro UV com 4 Dubon Ct Farmingdale NY 11735 866 230 7305 ...

Отзывы: