SERVOPRO CHROMA Gas Analyser
85
04400001A / Revision 7
Start the calibration:
9.
Enable the
Span Calibration
with the
Enable (F3)
button.
10.
In the
Span
column select
Yes
for the peaks that require calibration
11.
Select
All peaks
or
Selected peaks
by changing the slider state. When
Selected peaks
is selected, only peaks that have a
YES
in the
Select
column
of the grid
will be calibrated.
12.
Click on
Start
(F6)
button to start calibration
13.
After three cycles with consistent peak values (within 3% FS) press the
RE-
SPAN
button to accept the calibration. A pop-up box question will appear
Are
you sure you want to re-span?
Select
YES
to accept the calibration. The
Result
column will be updated with the value set in the
Span Gas
column,
and the
Response Factor
column will be re-calculated.
14.
When the
Span Calibration
is completed,
YES
is displayed in the
Span Done
column of the grid.
15.
Repeat the above if there are other peaks to be calibrated.
16.
Disable
Span
Calibration
with the
Disable (F3)
button (one button performs
dual functions Enable or Disable, depending on the active mode).
Figure 40 : Calibration-Span menu