![Spectris SERVOPRO Chroma Скачать руководство пользователя страница 135](http://html.mh-extra.com/html/spectris/servopro-chroma/servopro-chroma_operators-manual_1344053135.webp)
SERVOPRO CHROMA Gas Analyser
134
04400001A / Revision 7
12.11 FID Troubleshooting
Symptom
Problem cause
Possible remedy
Burner
doesn’t
ignite
Fuel gas emerging from
burner jet diluted with other
gases because fuel system is
insufficiently purged.
Purge fuel gas regulator.
Air and fuel flow not properly
set.
Adjust air and flow set points
to match the values on the
Test Results Sheet.
Glow plug damaged.
Contact Servomex.
Analyser not able to reach
starting flow set points
Verify all inlet pressures.
Analyser’s reading
is noisy
Air or fuel is contaminated.
Replace air or fuel cylinder.
Add hydrocarbon traps.
Purge tubing.
There is a leak on sample
line.
Leak test your sampling
system.
Flows are not stable
Verify that all gas pressures
are stable.
Verify that flow readings are
stable.
Re-validate PIDs for ovens
and flows
Tubing used to connect gases
to the SERVOPRO FID was
not properly purged.
Purge tubing
There is a dead volume on
one of the gas stream.
Verify plumbing diagram and
remove any dead volume.
Exhaust is blocked by water
Make sure the exhaust tubing
has a slope.
Loss of sensitivity
or no signal
Electrical elements of burner
partially short circuited by
combustion products.
Contact Servomex.
Polarisation spring is not in
contact with FID jet.
False flame
indication
Flame sensor may be
damaged.
Table 35 : FID Troubleshooting