3-2
Airnet® II 4-Channel Particle Sensors Operations Manual
Chapter 3
: Configuring for Ethernet
Configuration Defined
RS-232 Cable
A suitable cable (PMS-CD1995) is provided in your shipment. The wiring for this cable
is as follows:
Setting Configuration Addresses for Ethernet
Follow these steps to configure the Airnet II 4-Channel particle sensor to communicate with
Facility Net or any compatible host program by means of an Ethernet network.
To configure the Sensor:
1.
Connect the Airnet II 4-Channel particle sensor to a computer that has terminal
emulation software using the provided RS-232 cable.
2.
Start the terminal emulation software and proceed to the COM 1 parameters
dialog box.
3.
Set the following communication parameters:
Table 3-1
RS-232 pin outs (RJ-11 jack)
Pin 1
DSR
Pin 2
Transmit
Pin 3
Receive
Pin 4
RTS
Pin 5
CTS
Pin 6
GND
NOTE:
Pins 1, 4, and Pin 5 are shorted together.
CAUTION
As with any network-capable software, do not attempt to connect the Airnet II
particle sensor to a network without your network administrator’s cognizance
and explicit permission. The administrator will need to approve, and in most
cases will issue, addresses.
Table 3-2
Terminal emulator communications parameters
Baud Rate
9600
Data Bits
8
Parity
N
Stop Bits
1
Flow Control
Off
Parity Check
Off
Carrier Detect
Off